Literature DB >> 18436718

Expanding assay dynamics: a combined competitive and direct assay system for the quantification of proteins in multiplexed immunoassays.

Michael Hartmann1, Monika Schrenk, Anette Döttinger, Sarah Nagel, Johan Roeraade, Thomas O Joos, Markus F Templin.   

Abstract

BACKGROUND: The concurrent detection and quantification of analytes that vary widely in concentration present a principal problem in multiplexed assay systems. Combining competitive and sandwich immunoassays permits coverage of a wide concentration range, and both highly abundant molecules and analytes present in low concentration can be quantified within the same assay.
METHODS: The use of different fluorescence readout channels allows the parallel use of a competitive system and a sandwich configuration. The 2 generated assay signals are combined and used to calculate the amount of analyte. The measurement range can be adjusted by varying the competitor concentration, and an extension of the assay system's dynamic range is possible.
RESULTS: We implemented the method in a planar protein microarray-based autoimmune assay to detect autoantibodies against 13 autoantigens and to measure the concentration of a highly abundant protein, total human IgG, in one assay. Our results for autoantibody detection and IgG quantification agreed with results obtained with commercially available assays. The use of 2 readout channels in the protein microarray-based system reduced spot-to-spot variation and intraassay variation.
CONCLUSIONS: By combining a direct immunoassay with a competitive system, analytes present in widely varying concentrations can be quantified within a single multiplex assay. Introducing a second readout channel for analyte quantification is an effective tool for spot-to-spot normalization and helps to lower intraassay variation.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18436718     DOI: 10.1373/clinchem.2007.099812

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

1.  An automated integrated platform for rapid and sensitive multiplexed protein profiling using human saliva samples.

Authors:  Shuai Nie; W Hampton Henley; Scott E Miller; Huaibin Zhang; Kathryn M Mayer; Patty J Dennis; Emily A Oblath; Jean Pierre Alarie; Yue Wu; Frank G Oppenheim; Frédéric F Little; Ahmet Z Uluer; Peidong Wang; J Michael Ramsey; David R Walt
Journal:  Lab Chip       Date:  2014-03-21       Impact factor: 6.799

2.  Advantages of multiplex proteomics in clinical immunology: the case of rheumatoid arthritis: novel IgXPLEX™: planar microarray diagnosis.

Authors:  Peter Lea; Edward Keystone; Sasi Mudumba; Anthony Kahama; Shi-Fa Ding; Jennifer Hansen; Azar A Azad; Sihe Wang; Deborah Weber
Journal:  Clin Rev Allergy Immunol       Date:  2011-08       Impact factor: 8.667

Review 3.  Salivary diagnostics using a portable point-of-service platform: a review.

Authors:  Prarthana Khanna; David R Walt
Journal:  Clin Ther       Date:  2015-02-27       Impact factor: 3.393

Review 4.  Protein microarrays: high-throughput tools for proteomics.

Authors:  Oda Stoevesandt; Michael J Taussig; Mingyue He
Journal:  Expert Rev Proteomics       Date:  2009-04       Impact factor: 3.940

  4 in total