Literature DB >> 18433842

Comparison of the ecarin chromogenic assay and different aPTT assays for the measurement of argatroban concentrations in plasma from healthy individuals and from coagulation factor deficient patients.

Robert Siegmund1, Klas Boer, Katrin Poeschel, Gunter Wolf, Thomas Deufel, Michael Kiehntopf.   

Abstract

INTRODUCTION: The aim of this study was to assess the usefulness of different aPTT assays and the ecarin chromogenic assay reaction time (ECA) for measurement of argatroban concentration in plasma from healthy persons as well as in different patient subgroups.
METHODS: We spiked plasma samples from healthy individuals, patients under oral anticoagulation (OAT) or with liver dysfunction (LD) with increasing argatroban concentrations (0-2000 ng/ml) and performed 4 different aPTTs assays and the ECA.
RESULTS: Depending on argatroban concentrations aPTTs increased in a curvilinear fashion; in plasma from healthy individuals means of calculated argatroban concentration at 2-fold aPTT differed extensively depending on the aPTT reagent used (725 ng/ml to 1136 ng/ml) and were even more pronounced in plasma from coagulation factor deficient patients (460 ng/ml in patients with LD vs. 1172 ng/ml in patients with OAT), whereas ECA showed linear argatroban influence and reliable results in all subgroups.
CONCLUSIONS: Because of wide differences in aPTT measurements depending on the aPTT reagent used, interindividual variations and different clinical conditions the aPTT is not the method of choice for monitoring argatroban and the ECA should be preferred.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18433842     DOI: 10.1016/j.thromres.2008.02.013

Source DB:  PubMed          Journal:  Thromb Res        ISSN: 0049-3848            Impact factor:   3.944


  3 in total

1.  Evaluation of intravenous direct thrombin inhibitor monitoring tests: Correlation with plasma concentrations and clinical outcomes in hospitalized patients.

Authors:  Jacob T Beyer; Stuart E Lind; Sheila Fisher; Toby C Trujillo; Michael F Wempe; Tyree H Kiser
Journal:  J Thromb Thrombolysis       Date:  2020-02       Impact factor: 2.300

2.  Ecarin modified rotational thrombelastometry: a point-of-care applicable alternative to monitor the direct thrombin inhibitor argatroban.

Authors:  Eva Schaden; Andreas Schober; Stefan Hacker; Sibylle Kozek-Langenecker
Journal:  Wien Klin Wochenschr       Date:  2013-02-26       Impact factor: 1.704

Review 3.  Reducing harm associated with anticoagulation: practical considerations of argatroban therapy in heparin-induced thrombocytopenia.

Authors:  Marcie J Hursting; Joseph Soffer
Journal:  Drug Saf       Date:  2009       Impact factor: 5.606

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.