Dynact soft-tissue visualization using an angiographic C-arm system: initial clinical experience in the operating room.

Koreaki Irie1, Yuichi Murayama, Takayuki Saguchi, Toshihiro Ishibashi, Masaki Ebara, Hiroyuki Takao, Toshiaki Abe.   

Abstract

INTRODUCTION: DynaCT is a clinical application protocol to create computed tomographic (CT)-like images allowing soft-tissue visualization acquired from an angiography system. A cone beam three-dimensional CT reconstruction is produced from the acquisition of two-dimensional projection images by rotating the c-arm with x-ray source and image receptor around the patient. The purpose of this study is to evaluate the clinical efficacy of DynaCT in the operating room.
METHODS: DynaCT was performed in 100 patients undergoing cerebral or spinal interventional procedures in the new-concept operating room. Specially designed AXIOM Artis BA (conventional image intensifier system; 55 patients; Siemens Medical Solutions, Erlangen, Germany) and AXIOM Artis dBA (flat-panel detector; 45 patients; Siemens Medical Solutions) biplane neuroangiographic systems (Siemens Medical Solutions, Germany) were installed in the operating room. The volumetric data set from the AXIOM Artis systems were reconstructed immediately on the three-dimensional workstation in the operating room. We compared DynaCT images with the use of multidetector computed tomography.
RESULTS: DynaCT was performed successfully in all patients. High-contrast structures, such as bone, calcified lesions, and metallic materials, were visualized on DynaCT as well as in multidetector computed tomography for each group. Although contrast differentiation of soft tissue such as cerebral cortex, muscle, and hematoma on DynaCT were inferior to multidetector CT scans, the images were sufficiently used as intraoperative CT-like images. However, DynaCT images acquired from flat panel detector-based systems were found to be superior to those images acquired from image intensifier-based systems. Striking ring artifacts were exhibited and resulted in major limitations in the image intensifier group.
CONCLUSION: DynaCT has the potential to be used as a powerful tool for endovascular and neurosurgical procedures and will open new possibilities for neurosurgical management.

Entities:  

Mesh:

Year:  2008        PMID: 18424996     DOI: 10.1227/01.neu.0000317403.23713.92

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  8 in total

1.  Incidence and neurological outcomes of aneurysm rupture during interventional neuroradiology procedures in a hybrid operating suite.

Authors:  Kentaro Yamakawa; Shuya Kiyama; Yuichi Murayama; Shoichi Uezono
Journal:  J Anesth       Date:  2012-02-24       Impact factor: 2.078

2.  Comparison of air kerma between C-arm CT and 64-multidetector-row CT using a phantom.

Authors:  Seiki Hosokawa; Nobuyuki Kawai; Morio Sato; Hiroki Minamiguchi; Motoki Nakai; Kazuhiro Murotani; Tadayoshi Nishioku; Shintaro Shirai; Tetsuo Sonomura
Journal:  Jpn J Radiol       Date:  2014-02-20       Impact factor: 2.374

3.  Initial evaluation of image performance of a 3-D x-ray system: phantom-based comparison of 3-D tomography with conventional computed tomography.

Authors:  Robyn Melanie Benz; Meritxell Alzamora Garcia; Felix Amsler; Johannes Voigt; Andreas Fieselmann; Anna Lucja Falkowski; Bram Stieltjes; Anna Hirschmann
Journal:  J Med Imaging (Bellingham)       Date:  2018-03-06

4.  Lens dose reduction with a bismuth shield in neuro cone-beam computed tomography: an investigation on optimum shield device placement conditions.

Authors:  Satoru Kawauchi; Koichi Chida; Yusuke Hamada; Wataro Tsuruta
Journal:  Radiol Phys Technol       Date:  2021-11-18

5.  Use of cone-beam computed tomography angiography in planning for gamma knife radiosurgery for arteriovenous malformations: a case series and early report.

Authors:  Mina G Safain; Jason P Rahal; Ami Raval; Mark J Rivard; John E Mignano; Julian K Wu; Adel M Malek
Journal:  Neurosurgery       Date:  2014-06       Impact factor: 4.654

6.  3D C-arm conebeam CT angiography as an adjunct in the precise anatomic characterization of spinal dural arteriovenous fistulas.

Authors:  T D Aadland; K R Thielen; T J Kaufmann; J M Morris; G Lanzino; D F Kallmes; B A Schueler; H Cloft
Journal:  AJNR Am J Neuroradiol       Date:  2009-10-22       Impact factor: 3.825

7.  High-resolution cone beam CT for evaluation of vascular channel in intracranial partial thrombosed aneurysm.

Authors:  Kazunori Shintai; Noriaki Matsubara; Takashi Izumi
Journal:  Nagoya J Med Sci       Date:  2018-05       Impact factor: 1.131

8.  Usefulness of preoperative cone beam computed tomography and intraoperative digital subtraction angiography for dural arteriovenous fistula at craniocervical junction: Technical case report.

Authors:  Nobuhide Hayashi; Nagatsuki Tomura; Hideo Okada; Takahiro Sasaki; Eisaku Tsuji; Hiroki Enomoto; Toshikazu Kuwata
Journal:  Surg Neurol Int       Date:  2019-01-18
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.