Literature DB >> 18409670

[Adequacy and reliability of orthoanalyzer Ergovision for job-fitness evaluation].

B Totaro1, R Assini, D Consonni, C Guzzi, P Troiano, R Dal Pozzo, M D'Orso, A Bergamaschi, B Piccoli.   

Abstract

The aim of this research was to evaluate the Ergovision Screener (ES) accuracy e validity by a confrontation with the conventional ophthalmic check (OC), for the medical evaluation of job fitness. A population of 100 VDU operators was considered. Each subject underwent randomly both the ES and the ophthalmic check visit. Several test carried out by the Ergovision Screener were not consistent with the conventional ophthalmic check. In a number of cases, high false positive ratio have been found, which could lead to unnecessary further examinations. For all these reasons we believe that the ES is not an appropriate instrument for the medical evaluation of job fitness.

Entities:  

Mesh:

Year:  2007        PMID: 18409670

Source DB:  PubMed          Journal:  G Ital Med Lav Ergon        ISSN: 1592-7830


  1 in total

1.  Diagnosis accuracy of two vision screeners for visual health surveillance of workers who use video display terminals.

Authors:  María-José Molina-Torres; María-Del-Mar Seguí Crespo; Ana Tauste Francés; Blanca Lumbreras Lacarra; Elena Ronda-Pérez
Journal:  J Occup Health       Date:  2016-08-04       Impact factor: 2.708

  1 in total