Hua Xie1, Yashodhan S Khajanchee, Brian S Shaffer.
Abstract
BACKGROUND AND OBJECTIVES: This study was to evaluate the feasibility of using a novel chitosan hemostatic dressing to control hemorrhage and urinary leakage by sealing off the parenchymal wound following LPN.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18402734 PMCID: PMC3016030
Source DB: PubMed Journal: JSLS ISSN: 1086-8089 Impact factor: 2.172
Operation Data of Laparoscopic Partial Nephrectomy With Chitosan Dressings
| Polar Resection (N = 13) | Wedge Resection (N = 5) | P Value | |
|---|---|---|---|
| Resection Weight, g | 25 ± 4 | 19 ± 6 | 0.19 |
| Number of Applications (range) | 1.5 ± 0.7 (1 to 3) | 2.4 ± 1.1 (1 to 4) | 0.09 |
| Application Time | 14 ± 9 | 27 ± 25 | 0.29 |
| Estimated Blood Loss, mL, (range) | 51 ± 74 (10 to 250) | 121 ± 160 (10 to 400) | 0.40 |
| Changes of Hemostasis Score (range) | 3.2 ± 0.8 (2 to 4) | 3 ± 1.7 (0 to 4) | 0.50 |
| Urine Leakage (%, n) | (15%, 2/13) | (40%, 2/5) | 0.32 |
Application time represents the time required for successful deployment of the dressing and achieving secure adhesion of the dressing to the renal parenchymal injury.
Summary of Laparoscopic Partial Nephrectomy in Select Literature
| Authors | No. Pts | Mean Lesion Size (cm) | Mean OR Time (min) | Mean EBL (mL) | Convert to Open (n) | Complications (n, %) | Hemostatic Technique | |
|---|---|---|---|---|---|---|---|---|
| Leak | Bleed | |||||||
| Rassweiler et al, 2000[ | 53 | 2.4 | 191 | 725 | 4 | 5, 9.4 | 5, 9.4 | Bipolar, fibrin glue coated cellulose,Gelatin resorcinol formaldehyde glue |
| Kim et al, 2003[ | 79 | 2.5 | 182 | 391 | 1 | 2, 2.5 | 5, 6.3 | Electrocautery, Polyglactin bolster, suture |
| Johnston et al, 2005[ | 100 | 2.5 | 191 | 358 | 3 | 13, 13.0 | 9, 9.0 | FB, suture, bipolar, ABC, FloSeal®,Gelfoam® bolster |
| Ramani et al, 2005[ | 200 | 2.9 | 199 | 247 | 2 | 4, 2.0 | 28, 14.0 | Monopolar, Surgicel® bolster, suture |
| Gill et al, 2005[ | 63 | 2.5 | 218 | 106 | - | 1, 1.6 | 2, 3.2 | Monopolar, Surgicel® bolster, FloSeal ®, suture |
| Weld at al, 2006[ | 60 | 2.4 | 179 | 226 | - | 5, 8.3 | 2, 3.3 | FB, Harmonic, ABC, Bipolar, Surgicel® Bolster, FloSeal® and suture |
EBL = estimated blood loss; FB = fibrin glue; ABC = argon beam coagulation; OR = operating room.