Sensitivity evaluation of positron emission tomographs.

H W Reist1, J Vacher.   

Abstract

Applying the recommended guidelines of the Instrumentation Group of the EEC Concerted Action, the sensitivity of an ECAT 933/4-16 scanner was determined. Several filters were used to reconstruct the images. On the images regions of interest (ROIs) were defined, and the ROI-counts were compared to the specific activity in the phantom. Whereas the uncorrected sensitivity was not influenced by the reconstruction filters, the sensitivity corrected for the scatter fraction appeared to be dependent on the reconstruction filter. Images reconstructed with a non-smoothing filter (Ramp filter, Shepp filter) yielded a higher sensitivity due to the lower reconstructed scatter fraction.

Mesh:

Year:  1991        PMID: 1839844

Source DB:  PubMed          Journal:  Med Prog Technol        ISSN: 0047-6552