Literature DB >> 18382591

Charge-coupled device spectrograph for direct solar irradiance and sky radiance measurements.

Natalia Kouremeti1, Alkiviadis Bais, Stelios Kazadzis, Mario Blumthaler, Rainer Schmitt.   

Abstract

The characterization of a charged-coupled device (CCD) spectrograph developed at the Laboratory of Atmospheric Physics, Thessaloniki is presented. The absolute sensitivity of the instrument for direct irradiance and sky radiance measurements was determined, respectively, with an uncertainty of 4.4% and 6.6% in the UV-B, and 3% and 6% in the UV-A, visible and near-infrared (NIR) wavelength ranges. The overall uncertainty associated with the direct irradiance and the sky radiance measurements is, respectively, of the order of 5% and 7% in the UV-B, increasing to 10% for low signals [e.g., at solar zenith angles (SZAs) larger than 70 degrees ], and 4% and 6% in the UV-A, visible, and NIR. Direct solar spectral irradiance measurements from an independently calibrated spectroradiometer (Bentham DTM 300) were compared with the corresponding CCD measurements. Their agreement in the wavelength range of 310-500nm is within 0.5% +/- 1.1% (for SZA between 20 degrees and 70 degrees ). Aerosol optical depth (AOD) derived by the two instruments using direct Sun spectra and by a collocated Cimel sunphotometer [Aerosol Robotic network (AERONET)] agree to within 0.02 +/- 0.02 in the range of 315-870 nm. Significant correlation coefficients with a maximum of 0.99 in the range of 340-360 nm and a minimum of 0.90 at 870 nm were found between synchronous AOD measurements with the Bentham and the Cimel instruments.

Entities:  

Year:  2008        PMID: 18382591     DOI: 10.1364/ao.47.001594

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  The Solar Spectrum in the Atacama Desert.

Authors:  R R Cordero; A Damiani; G Seckmeyer; J Jorquera; M Caballero; P Rowe; J Ferrer; R Mubarak; J Carrasco; R Rondanelli; M Matus; D Laroze
Journal:  Sci Rep       Date:  2016-03-02       Impact factor: 4.379

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.