Literature DB >> 18375307

Treatment of infolding related to the gore TAG thoracic endoprosthesis.

Daniel A Leung1, Ivan Davis, Gundars Katlaps, Jaime Tisnado, Malcolm K Sydnor, Daniel J Komorowski, Derek Brinster.   

Abstract

The present report describes three cases of thoracic aortic endograft infolding or collapse involving the Gore TAG system. The cases include a penetrating aortic injury, a blunt aortic injury, and a symptomatic type B dissection. In the first case, infolding occurred in a delayed fashion after a normal-appearing 3-month follow-up computed tomographic angiogram. In the other two cases, infolding occurred during the immediate postoperative phase. One of the patients underwent explantation and surgical repair. The other two underwent endovascular repair of the infolded endograft by placement of a balloon-expandable stent in one case and a self-expanding stent in the other.

Entities:  

Mesh:

Year:  2008        PMID: 18375307     DOI: 10.1016/j.jvir.2007.12.453

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Delayed endovascular treatment of descending aorta stent graft collapse in a patient treated for post- traumatic aortic rupture: a case report.

Authors:  Giovanni Nano; Daniela Mazzaccaro; Giovanni Malacrida; Maria Teresa Occhiuto; Silvia Stegher; Domenico G Tealdi
Journal:  J Cardiothorac Surg       Date:  2011-05-24       Impact factor: 1.637

2.  Infolding of fenestrated endovascular stent graft.

Authors:  Jason G E Zelt; Prasad Jetty; Adnan Hadziomerovic; Sudhir Nagpal
Journal:  J Vasc Surg Cases Innov Tech       Date:  2017-07-20
  2 in total