Dynamic contour tonometry in keratoconus and postkeratoplasty eyes.

Alexander Meyenberg1, Milko E Iliev, Raphael Eschmann, Beatrice E Frueh.   

Abstract

PURPOSE: To compare the performance of dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in measuring intraocular pressure in eyes with irregular corneas.
METHODS: GAT and DCT measures were taken in 30 keratoconus and 29 postkeratoplasty eyes of 35 patients after pachymetry and corneal topography. Regression and correlation analyses were performed between both tonometry methods and between tonometry methods and corneal parameters. Bland-Altman plots were constructed.
RESULTS: DCT values were significantly higher than GAT values in both study groups: +4.1 +/- 2.3 mm Hg (mean +/- SD) in keratoconus and +3.1 +/- 2.5 mm Hg after keratoplasty. In contrast to DCT, GAT values were significantly higher in postkeratoplasty eyes than in keratoconus. The correlation between the 2 tonometry methods was moderate in keratoconus (Kendall correlation coefficient, tau = 0.34) as well in postkeratoplasty eyes (tau = 0.66). The +/-1.96 SD span of the DCT-GAT differences showed a considerable range: -0.42 to +8.70 mm Hg in keratoconus and -1.87 to +7.98 mm Hg in postkeratoplasty eyes. In the keratoconus group, neither DCT nor GAT correlated significantly with any of the corneal parameters. In the postkeratoplasty group, both DCT and GAT measures showed a moderate positive correlation with corneal steepness, but only DCT had a significant negative correlation with the central corneal thickness (tau = -0.33).
CONCLUSIONS: DCT measured significantly higher intraocular pressures than GAT in keratoconus and postkeratoplasty eyes. DCT and GAT measures varied considerably, and DCT was not less dependent on biomechanical properties of irregular corneas than GAT.

Entities:  

Mesh:

Year:  2008        PMID: 18362658     DOI: 10.1097/ICO.0b013e31815dcf97

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  5 in total

1.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

2.  Assessment of intraocular pressure measured by Reichert Ocular Response Analyzer, Goldmann Applanation Tonometry, and Dynamic Contour Tonometry in healthy individuals.

Authors:  Ping-Bo Ouyang; Cong-Yi Li; Xiao-Hua Zhu; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-02-18       Impact factor: 1.779

3.  Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus.

Authors:  Fatih Özcura; Nilgün Yıldırım; Emre Tambova; Afsun Şahin
Journal:  J Optom       Date:  2016-07-09

4.  Utility of Goldmann applanation tonometry for monitoring intraocular pressure in glaucoma patients with a history of laser refractive surgery.

Authors:  Sang Yeop Lee; Hyoung Won Bae; Hee Jung Kwon; Gong Je Seong; Chan Yun Kim
Journal:  PLoS One       Date:  2018-02-05       Impact factor: 3.240

Review 5.  Intraocular Pressure Measurement after Penetrating Keratoplasty.

Authors:  Otilia-Maria Dumitrescu; Sinziana Istrate; Mioara-Laura Macovei; Alina Gabriela Gheorghe
Journal:  Diagnostics (Basel)       Date:  2022-01-19
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.