A mechanical comparison between conventional and modified angular plates for proximal humeral fractures.

Eduardo F Carrera1, Fabio A Nicolao, Nicola Archetti Netto, Renato L Carvalho, Fernando B Dos Reis, Enrico José Giordani.   

Abstract

The objective of this study is to present a modified angular blade plate for fixing 2-part and even 3-part fractures of the proximal humerus, as well as the results of the comparative mechanical test between the conventional angular blade plate and this new modified plate. The plates were tested in flexion and rotational trials in a wooden model that simulated a 2-part humeral fracture of the proximal extremity. The results (mean +/- SD) of bending strength and stiffness obtained after testing showed findings of 601 +/- 349 N and 0.5 +/- 0.2 N/mm, respectively, for the conventional plate and 4005 +/- 164 N and 3.9 +/- 0.7 N/mm, respectively, for the modified plate. The torsional stiffness test showed findings of 1.26 +/- 0.09 KN.mm degrees for the conventional plate and 1.74 +/- 0.21 KN.mm degrees for the modified plate. The test of torsional moment showed findings of 57.0 +/- 7.6 KN.mm for the conventional plate and 115.2 +/- 9.3 KN.mm for the modified plate. The test of angular displacement at the torsional moment showed findings of 50.8 degrees +/- 7.2 degrees for the conventional plate and 70.2 degrees +/- 2.6 degrees for the modified plate. The results of the mechanical trials of flexion and rotation were superior for the modified angular blade plate compared with the conventional angular blade plate.

Entities:  

Mesh:

Year:  2008        PMID: 18359646     DOI: 10.1016/j.jse.2008.02.003

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  4 in total

1.  Biomechanics of four techniques for fixation of the four-part humeral head fracture.

Authors:  Elpídio da Graça; Rodrigo Okubo; Antônio Carlos Shimano; Nilton Mazzer; Cláudio Henrique Barbieri
Journal:  Acta Ortop Bras       Date:  2013-01       Impact factor: 0.513

2.  Comparing the In Vitro Stiffness of Straight-DCP, Wave-DCP, and LCP Bone Plates for Femoral Osteosynthesis.

Authors:  José Ricardo Lenzi Mariolani; William Dias Belangero
Journal:  ISRN Orthop       Date:  2013-02-26

Review 3.  Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review.

Authors:  Ali Jabran; Chris Peach; Lei Ren
Journal:  Biomed Eng Online       Date:  2018-04-27       Impact factor: 2.819

Review 4.  An overview of de novo bone generation in animal models.

Authors:  Takashi Taguchi; Mandi J Lopez
Journal:  J Orthop Res       Date:  2020-09-23       Impact factor: 3.494