Literature DB >> 18355817

Accuracy of the size estimation in wireless capsule endoscopy: calibrating the M2A PillCam (with video).

Florian Graepler1, Manuel Wolter, Reinhard Vonthein, Michael Gregor.   

Abstract

OBJECTIVE: To describe the accuracy and precision of a size estimation in wireless capsule endoscopy (WCE) and to develop a simple device that facilitates learning of an exact size estimation in WCE.
DESIGN: A prospective study with an animal model. Seventy-five medical students were compared with 21 physicians not performing WCEs and 21 experts actively performing WCEs.
INTERVENTIONS: Video sequences were recorded with a PillCam wireless endoscope in a porcine small-bowel model after implantation of "lesions" of defined sizes. Volunteers estimated the diameter of these "lesions" with and without the use of a calibrator device, which was a picture from a black-and-white grid taken with the PillCam in direct contact with the transparent dome.
RESULTS: The students overestimated the diameters of the lesions by 26%. The physicians and experts underestimated the diameters by 32% and 31%, respectively. With the help of the calibrator device, the students underestimated the diameters by 15%. Physicians underestimated the diameters by 21%, whereas experts still underestimated the diameters by 35%. The interindividual log-scale standard deviation (SD) of the estimated diameters from the true diameter decreased during training in all groups, whereas intraindividual SDs decreased in students and increased in experts. LIMITATION: An animal model.
CONCLUSIONS: All investigators agree better when a calibrating device was offered. Experts who performed more than 400 WCEs tended to be more precise.

Entities:  

Mesh:

Year:  2008        PMID: 18355817     DOI: 10.1016/j.gie.2007.10.060

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  3 in total

Review 1.  New vision in video capsule endoscopy: current status and future directions.

Authors:  Laurel R Fisher; William L Hasler
Journal:  Nat Rev Gastroenterol Hepatol       Date:  2012-05-08       Impact factor: 46.802

2.  Training in Capsule Endoscopy: Are We Lagging behind?

Authors:  Reena Sidhu; Mark E McAlindon; Carolyn Davison; Simon Panter; Olaf Humbla; Martin Keuchel
Journal:  Gastroenterol Res Pract       Date:  2012-04-11       Impact factor: 2.260

3.  A Simple Evaluation Tool (ET-CET) Indicates Increase of Diagnostic Skills From Small Bowel Capsule Endoscopy Training Courses: A Prospective Observational European Multicenter Study.

Authors:  J G Albert; O Humbla; M E McAlindon; C Davison; U Seitz; C Fraser; F Hagenmüller; E Noetzel; C Spada; M E Riccioni; J Barnert; N Filmann; M Keuchel
Journal:  Medicine (Baltimore)       Date:  2015-10       Impact factor: 1.817

  3 in total