| Literature DB >> 18349901 |
A S Bakulin1, S M Durbin, T Jach, J Pedulla.
Abstract
An improved x-ray microscope with a fully electronic CCD detector system has been constructed that allows improved laboratory-based microstructural investigations of materials with hard x rays. It uses the Kirkpatrick-Baez multilayer mirror design to form an image that has a demonstrated resolution of 4 microm at 8 keV (Cu K(alpha) radiation). This microscope performs well with standard sealed-tube laboratory x-ray sources, producing digital images with 20-s exposure times for a 5-microm Au grid (a thickness of two absorption lengths).Entities:
Year: 2000 PMID: 18349901 DOI: 10.1364/ao.39.003333
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980