AquaLase versus NeoSoniX--a comparison study.

Nada Jiraskova1, Pavel Rozsival, Jana Kadlecova, Jana Nekolova, Jana Pozlerova, Zlatica Dubravska.   

Abstract

AIMS: To compare the metrics and surgical outcome when using Infiniti AquaLase and NeoSoniX cataract removal modalities.
METHODS: This prospective clinical study involved 50 patients with bilateral cataracts and lens removal using AquaLase in the right eye and NeoSoniX in the left eye. Best corrected visual acuity (BCVA), endothelial cell density and pachymetry were evaluted pre- and postoperatively. Statistical analysis was performed using the Wilcoxon Signed- Rank Test.
RESULTS: Preoperative mean pachymetry was 569 +/- 31 mu in the right eye (RE) and 560 +/- 37 mu in the left eye (LE), mean endothelial cell density 2744 +/- 418 cells/mm(2) (RE) and 2730 +/- 472 cells/mm(2) (LE). One week after operation pachymetry was 576 +/- 52 mu (RE) and 583 +/- 72 mu (LE) and endothelial cell density 2388 +/- 586 cells/mm(2) (RE) and 2463 +/- 615 cells/mm(2) (LE). One month after surgery pachymetry was 556 +/- 43 mu (RE) and 559 +/- 44 mu (LE) and endothelial cell density 2368 +/- 52 cells/mm(2) (RE) and 2495 +/- 548 cells/mm(2) (LE). BCVA improved in all eyes and was 0.8 or better on the first postoperative day.
CONCLUSIONS: Both the NeosoniX and AquaLase minimize intraoperative damage to ocular structures.

Entities:  

Mesh:

Year:  2007        PMID: 18345270     DOI: 10.5507/bp.2007.053

Source DB:  PubMed          Journal:  Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub        ISSN: 1213-8118            Impact factor:   1.245


  1 in total

Review 1.  Liquefaction for cataract extraction.

Authors:  Georgios Labiris; Aspasia Toli; Damaskini Polychroni; Maria Gkika; Dimitrios Angelonias; Vassilios P Kozobolis
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.