Dosimetry quality audit of high energy photon beams in greek radiotherapy centers.

Constantine J Hourdakis1, A Boziari.   

Abstract

BACKGROUND AND
PURPOSE: Dosimetry quality audits and intercomparisons in radiotherapy centers is a useful tool in order to enhance the confidence for an accurate therapy and to explore and dissolve discrepancies in dose delivery. This is the first national comprehensive study that has been carried out in Greece. During 2002--2006 the Greek Atomic Energy Commission performed a dosimetry quality audit of high energy external photon beams in all (23) Greek radiotherapy centers, where 31 linacs and 13 Co-60 teletherapy units were assessed in terms of their mechanical performance characteristics and relative and absolute dosimetry.
MATERIALS AND METHODS: The quality audit in dosimetry of external photon beams took place by means of on-site visits, where certain parameters of the photon beams were measured, calculated and assessed according to a specific protocol and the IAEA TRS 398 dosimetry code of practice. In each radiotherapy unit (Linac or Co-60), certain functional parameters were measured and the results were compared to tolerance values and limits. Doses in water under reference and non reference conditions were measured and compared to the stated values. Also, the treatment planning systems (TPS) were evaluated with respect to irradiation time calculations.
RESULTS: The results of the mechanical tests, dosimetry measurements and TPS evaluation have been presented in this work and discussed in detail. This study showed that Co-60 units had worse performance mechanical characteristics than linacs. 28% of all irradiation units (23% of linacs and 42% of Co-60 units) exceeded the acceptance limit at least in one mechanical parameter. Dosimetry accuracy was much worse in Co60 units than in linacs. 61% of the Co60 units exhibited deviations outside +/-3% and 31% outside +/-5%. The relevant percentages for the linacs were 24% and 7% respectively. The results were grouped for each hospital and the sources of errors (functional and human) have been investigated and discussed in details.
CONCLUSION: This quality audit proved to be a useful tool for the improvement of quality in radiotherapy. It succeeded to disseminate the IAEA TRS-398 protocol in nearly all radiotherapy centers achieving homogenization and consistency of dosimetry within the country. Also, it detected discrepancies in dosimetry and provided guidance and recommendations to eliminate sources of errors. Finally, it proved that quality assurance programs, periodic quality control tests, maintenance and service play an important role for achieving accuracy and safe operation in radiotherapy.

Entities:  

Mesh:

Year:  2008        PMID: 18343514     DOI: 10.1016/j.radonc.2008.01.023

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  2 in total

1.  Assessment of national dosimetry quality audits results for teletherapy machines from 1989 to 2015.

Authors:  Wazir Muhammad; Asad Ullah; Khalid Mahmood
Journal:  J Appl Clin Med Phys       Date:  2016       Impact factor: 2.102

2.  An analysis of the regulatory program of quality audits in radiotherapy in Brazil from 1995 to 2007.

Authors:  Eduardo de Paiva; Luiz A R da Rosa; Ricardo R A Brito; Lidia V de Sá; Ana C M Dovales; Delano V S Batista; Ricardo A Giannoni; Alexandre F Velasco
Journal:  J Appl Clin Med Phys       Date:  2011-01-30       Impact factor: 2.243

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.