Plain radiographs are inadequate to identify overlengthening with a radial head prosthesis.

Heidi C Shors1, Caitlin Gannon, Mark C Miller, Christopher C Schmidt, Mark E Baratz.   

Abstract

PURPOSE: A correctly sized radial head implant helps restore more normal elbow motion and load distribution and allows the collateral ligaments to heal in an anatomic position. No single method of measurement has been agreed on. We hypothesized that plain radiographs could be used to evaluate the appropriate length of a radial head prosthesis when we simulated different patterns of ligament disruption.
METHODS: Osteotomies of the medial and lateral condyles were created in 6 cadaver specimens to simulate 4 conditions: ulnar collateral ligament (UCL) and lateral ulnar collateral ligament (LUCL) intact, UCL intact, LUCL intact, and UCL and LUCL disrupted. Radial heads were resected and replaced with correctly or incorrectly sized monopolar radial head implants. Radiographs were taken after simulated ligament injury with a radial head prosthesis that was either 2 mm too short, the appropriate length, or too long by either 2 or 4 mm. Measurements were recorded between radiographic markers to determine ulnohumeral joint displacement. A 4 x 4 x 2 repeated-measures analysis of variance was performed, as well as post hoc analyses using Tukey's honest significant difference test, to determine statistically significant changes in our radiographic measurements.
RESULTS: Results showed that disruption of at least 1 ligament alters the ulnohumeral joint separation on the lateral but not on the anteroposterior radiograph. Overlengthening the radial head did not result in a statistically significant change in ulnohumeral space.
CONCLUSIONS: Replicating radial length is an important but sometimes difficult step during radial head replacement. Standard radiographic assessment of the ulnohumeral joint will not demonstrate marked changes in the ulnohumeral space when the implant is underlengthened by 2 mm or overlengthened by as much as 4 mm, as long as the joint remains located. Disruption of at least 1 collateral ligament alters the ulnohumeral joint separation on the lateral but not on the anteroposterior radiograph.

Mesh:

Year:  2008        PMID: 18343288     DOI: 10.1016/j.jhsa.2007.12.004

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  7 in total

1.  Can ulnar variance be used to detect overstuffing after radial head arthroplasty?

Authors:  Jun-Gyu Moon; Jin-Ho Hong; Nitin Bither; Won-Yong Shon
Journal:  Clin Orthop Relat Res       Date:  2013-09-13       Impact factor: 4.176

2.  Simple overlay device for determining radial head and neck height.

Authors:  Jun-Gyu Moon; Richard D Southgate; James S Fitzsimmons; Shawn W O'Driscoll
Journal:  Skeletal Radiol       Date:  2010-03-01       Impact factor: 2.199

3.  Comminuted radial head fractures treated with pyrocarbon prosthetic replacement.

Authors:  Claudia Lamas; Juan Castellanos; Ignacio Proubasta; Enrique Dominguez
Journal:  Hand (N Y)       Date:  2010-06-15

Review 4.  Radial head prosthesis: surgical tips and tricks.

Authors:  A Marinelli; E Guerra; A Ritali; M Cavallo; R Rotini
Journal:  Musculoskelet Surg       Date:  2017-10-09

5.  Radial head replacement with unipolar and bipolar SBi system: a clinical and radiographic analysis after a 2-year mean follow-up.

Authors:  Roberto Rotini; Alessandro Marinelli; Enrico Guerra; Graziano Bettelli; Michele Cavaciocchi
Journal:  Musculoskelet Surg       Date:  2012-04-18

Review 6.  Complications of radial head prostheses.

Authors:  Stéphanie Delclaux; Julie Lebon; Amélie Faraud; Julien Toulemonde; Nicolas Bonnevialle; Bertrand Coulet; Pierre Mansat
Journal:  Int Orthop       Date:  2015-02-06       Impact factor: 3.075

7.  Is the distal radioulnar variance useful for identification of radial head prosthesis overlengthening? A cadaver study.

Authors:  Kilian Wegmann; Wei Zhang; Volker Strauss; Christian Ries; Lars P Müller; Klaus J Burkhart
Journal:  Eur J Med Res       Date:  2016-03-21       Impact factor: 2.175

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.