Literature DB >> 18341953

Membrane degradation of covered stents in the upper gastrointestinal tract: frequency and clinical significance.

Jin Hyoung Kim1, Ho-Young Song, Ji Hoon Shin, Hwoon-Yong Jung, Sung-Bae Kim, Jong-Hoon Kim, Seung-Il Park.   

Abstract

PURPOSE: To evaluate the frequency, clinical significance, and predictive factors of membrane degradation of covered stents in the upper gastrointestinal (UGI) tract.
MATERIALS AND METHODS: From 1996 to 2006, 166 stents were removed from 151 patients. Indications for stent removal included temporary stent placement for esophageal cancer before radiation therapy or radiation-chemotherapy (n = 53), stent migration (n = 40), temporary stent placement for a benign UGI stricture (n = 32), pain (n = 21), recurrent obstruction (n = 16), incomplete stent expansion (n = 1), recurrent transesophageal fistula (n = 2), and aspiration (n = 1). Removed stents were examined to evaluate possible causes of membrane degradation. Multivariate analysis was performed to determine the predictive factors of membrane degradation.
RESULTS: Degradation of the covering membrane occurred in 14 of the 166 stents (8%). Recurrent obstruction due to tumor ingrowth (n = 7) and reopening of the transesophageal fistula (n = 1) through the degraded membrane occurred in eight stents (5%). At multivariate logistic regression analysis, the type of covering membrane (P = .018), stricture location (P = .006), and duration of stent placement (P = .002) were significantly associated with degradation of the covering membrane.
CONCLUSIONS: Degradation of the covering membrane after stent placement in the UGI tract is not an uncommon event, and recurrent obstruction or fistula can occur through the degraded membrane. A polyurethane membrane is not biostable and dissolves over time, particularly in the gastroduodenal area.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18341953     DOI: 10.1016/j.jvir.2007.09.023

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  9 in total

1.  How to design the optimal self-expandable oesophageal metallic stents: 22 years of experience in 645 patients with malignant strictures.

Authors:  Han Kyu Na; Ho-Young Song; Jin Hyoung Kim; Jung-Hoon Park; Min Kyoung Kang; Jongjin Lee; Se Jin Oh
Journal:  Eur Radiol       Date:  2012-09-27       Impact factor: 5.315

2.  Malignant gastric outlet obstructions: treatment with self-expandable metallic stents.

Authors:  Jin Hyoung Kim; Ho-Young Song; Ji Hoon Shin
Journal:  Gut Liver       Date:  2010-09-10       Impact factor: 4.519

3.  Endoscopic ultrasound-guided choledochoduodenostomies with fully covered self-expandable metallic stents.

Authors:  Tae Jun Song; Yil Sik Hyun; Sang Soo Lee; Do Hyun Park; Dong Wan Seo; Sung Koo Lee; Myung-Hwan Kim
Journal:  World J Gastroenterol       Date:  2012-08-28       Impact factor: 5.742

4.  Comparison of uncovered and covered stents for the treatment of malignant duodenal obstruction caused by pancreaticobiliary cancer.

Authors:  Sang Myung Woo; Dong Hwan Kim; Woo Jin Lee; Kyung Woo Park; Sang-Jae Park; Sung-Sik Han; Tae Hyun Kim; Young Hwan Koh; Hyun Bum Kim; Eun Kyung Hong
Journal:  Surg Endosc       Date:  2013-01-04       Impact factor: 4.584

5.  Efficacy of a metallic stent covered with a paclitaxel-incorporated membrane versus a covered metal stent for malignant biliary obstruction: a prospective comparative study.

Authors:  Sung Ill Jang; Jie-Hyun Kim; Jung Whan You; Kwangwon Rhee; Se Joon Lee; Ho Gak Kim; Jimin Han; Im Hee Shin; Sang-Heum Park; Dong Ki Lee
Journal:  Dig Dis Sci       Date:  2012-11-21       Impact factor: 3.199

6.  Efficacy of a Multiplex Paclitaxel Emission Stent Using a Pluronic® Mixture Membrane versus a Covered Metal Stent in Malignant Biliary Obstruction: A Prospective Randomized Comparative Study.

Authors:  Sung Ill Jang; Se Joon Lee; Seok Jeong; Don Haeng Lee; Myung-Hwan Kim; Hong Jin Yoon; Dong Ki Lee
Journal:  Gut Liver       Date:  2017-07-15       Impact factor: 4.519

7.  Histological Changes in the Bile Duct after Long-Term Placement of a Fully Covered Self-Expandable Metal Stent within a Common Bile Duct: A Canine Study.

Authors:  Sang Soo Lee; Tae Jun Song; Mee Joo; Do Hyun Park; Dong Wan Seo; Sung Koo Lee; Myung-Hwan Kim
Journal:  Clin Endosc       Date:  2014-01-24

Review 8.  Endoluminal occlusion devices: technology update.

Authors:  Tobias Zander; Samantha Medina; Guillermo Montes; Lourdes Nuñez-Atahualpa; Michel Valdes; Manuel Maynar
Journal:  Med Devices (Auckl)       Date:  2014-12-01

9.  Gemcitabine-incorporated polyurethane films for controlled release of an anticancer drug.

Authors:  Seong Hoon Choi; Il-Hoon Cho; Sangsoo Park
Journal:  Biomater Res       Date:  2019-11-20
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.