Performance of the Abbott RealTime HIV-1 assay for quantification of HIV-1 clades prevalent in China.

Sihong Xu1, Aijing Song, Xiuhua Li, Jingyun Li, Zuoyi Bao, Panyong Mao, Qing Zhao, Youchun Wang.   

Abstract

OBJECTIVE: To evaluate the performance of the Abbott RealTime HIV-1 assay for measuring viral loads of the prevalent HIV-1 clades in China. STUDY
DESIGN: Serially diluted samples, as well as 521 clinical samples from 213 untreated HIV-1-infected individuals, 56 HIV-1-infected patients receiving ART, 60 HCV- and 57 HBV-infected patients and 135 healthy blood donors, were tested with RealTime and EasyQ.
RESULTS: Both assays exhibited linearity coefficients of >0.98. RealTime and EasyQt detected HIV-1 RNA in 87.36% and 86.99% of 269 HIV-1-seropositive samples, respectively. The correlation coefficients between the two assays for quantifying HIV-1 clades B', BC and AE were 0.884, 0.813 and 0.881, respectively, and the mean differences between the two assays for clades B', BC and AE were -0.087, 0.314 and 0.559 log(10)IU/mL, respectively. The correlation coefficient between the two assays for measuring samples from patients receiving ART was 0.945 (mean difference, 0.085 log(10)IU/mL). No false-positive samples were found among the 60 HCV-infected patients, 57 HBV-infected patients and 135 healthy blood donors.
CONCLUSIONS: The Abbott RealTime HIV-1 assay shows good linearity and specificity. ART drugs and HIV-1 clades B' and BC do not affect the performance of the assay. Based on the comparison data, [corrected] clade AE may be more [corrected] readily detected by using this method [corrected]

Entities:  

Mesh:

Year:  2008        PMID: 18313353     DOI: 10.1016/j.jcv.2008.01.009

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  3 in total

1.  Performance of NucliSens HIV-1 EasyQ Version 2.0 compared with six commercially available quantitative nucleic acid assays for detection of HIV-1 in China.

Authors:  Sihong Xu; Aijing Song; Jianhui Nie; Xiuhua Li; Youchun Wang
Journal:  Mol Diagn Ther       Date:  2010-10-01       Impact factor: 4.074

2.  Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays.

Authors:  Lesley E Scott; Lara D Noble; Jackie Moloi; Linda Erasmus; Willem D F Venter; Wendy Stevens
Journal:  J Clin Microbiol       Date:  2009-05-06       Impact factor: 5.948

3.  Evaluation of the Abbott m2000rt RealTime HIV-1 assay with manual sample preparation compared with the ROCHE COBAS AmpliPrep/AMPLICOR HIV-1 MONITOR v1.5 using specimens from East Africa.

Authors:  John A Crump; Lesley E Scott; Emma Msuya; Anne B Morrissey; Ekyafyose E Kimaro; John F Shao; Wendy S Stevens
Journal:  J Virol Methods       Date:  2009-09-01       Impact factor: 2.014

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.