Defibrillation coil reversal: a rare cause of abnormal noise and inappropriate shocks.

Vinodh Jeevanantham1, Ethan Levine, Adam S Budzikowski, Abrar H Shah, James P Daubert.   

Abstract

We report a rare case of inadvertent reversal of the defibrillator shock coils during generator change which then resulted in both intermittent, recurrent "noise" on both the rate sensing and the shock electrogram leads and multiple inappropriate shocks.