Literature DB >> 18300913

Differentiating intrathecal catheter tip granulomas from normal magnetic resonance image distortion caused by metallic catheter tips.

Stephen T Magill1, Paul Wang, Jorge L Eller, Kim J Burchiel.   

Abstract

OBJECTIVE: To characterize the normal image distortion produced by metallic intrathecal catheter tips and define features that differentiate them from a catheter tip granuloma.
METHODS: Three Medtronic Inc. (Minneapolis, MN) intrathecal catheter models (M8709, M8711, and M8731) were placed in a water-filled phantom, which was secured with a spinal coil and scanned with a 1.5-T magnetic resonance imaging (MRI) scanner. MRI scans from two patients receiving intrathecal opiate therapy without a granuloma and one patient with a granuloma were analyzed for comparison.
RESULTS: All three catheter models produced a metallic susceptibility artifact, which distorted the local tissue image. The tip appeared as a round, dark void sandwiched between two bright bands in the sagittal plane. Axial views produced an oval-shaped, dark central void completely surrounded by a bright rim. On MRI scans, catheter tip diameter varied from 2.3 to 6.7 mm and tip length varied from 4.0 to 7.5 mm, which is a 74 to 480% increase over their actual size. The catheter tip appearance in the phantom correlated with the tip appearance in MRI scans from patients with intrathecal catheters. Images from a tip granuloma show a bright rim surrounding the catheter tip in the sagittal plane with or without contrast as well as postcontrast enhancement of the whole granuloma, and the catheter tip appeared greater than 5 mm in diameter and length.
CONCLUSION: When evaluating possible granuloma development around an intrathecal catheter tip, it is important to take into account the characteristic metallic susceptibility artifact and increased size of normal metallic catheter tips on MRI scans.

Entities:  

Mesh:

Year:  2008        PMID: 18300913     DOI: 10.1227/01.NEU.0000311083.96389.FB

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  2 in total

1.  A Patient with an Intradural Tumor: An Unexpected Finding.

Authors:  Devang Padalia; Allan R Escher; Nasrin N Aldawoodi; Neal Shah
Journal:  Cureus       Date:  2020-03-23

2.  Spinal cord compression secondary to intrathecal catheter-induced granuloma: a report of four cases.

Authors:  Paul M Arnold; Viraat Harsh; Seth M Oliphant
Journal:  Evid Based Spine Care J       Date:  2011-02