In vitro drug adsorption and plasma free hemoglobin levels associated with hollow fiber oxygenators in the extracorporeal life support (ECLS) circuit.

Thomas J Preston1, Ashley B Hodge, Jeffrey B Riley, Cheryl Leib-Sargel, Kathleen K Nicol.   

Abstract

The purpose of this study was to identify the percentage of fentanyl or morphine sulfate lost from adhesion to either the polyvinylchloride (PVC) tubing or the surface of two different hollow fiber oxygenators used in current extracorporeal life support circuits and to identify any difference in the plasma free hemoglobin (PFH) levels generated when using these oxygenator and/or drug combinations. For each drug examined, six simple circuits were assembled; for each drug, two circuits contained tubing without an oxygenator (control), two circuits contained the Jostra Quadrox D (Maquet Cardiopulmonary, AG Hirrlingen, Germany), and two circuits contained the Terumo Baby Rx (Terumo Cardiovascular Systems Corp., Ann Arbor, MI). Fentanyl or morphine sulfate was added to yield initial circuit concentrations equal to 1430 ng/mL, respectively. Throughout the 6-hour in vitro testing, samples to evaluate the drug and PFH levels were drawn at various time intervals. Significance in this study is defined as p < .05. Fentanyl's initial adsorption seems to be 80% in circuits without oxygenators, 86% in the circuits containing the Quadrox D oxygenator, and 83% in the circuits with the Baby Rx oxygenator. Morphine sulfate seems to be initially adsorbed at a rate of 40% in all circuits and does not seem to be adsorbed by either of the tested oxygenators. The PFH levels were significantly (p < .05) elevated in the fentanyl circuits. The type of oxygenator does not seem to play a significant role in drug adsorption. During this in vitro study, the majority of both drugs were lost to the PVC tubing. The type of oxygenator did not seem to significantly affect PFH. However, fentanyl in any combination or alone was associated with increased PFH levels.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18293808      PMCID: PMC4680688     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  9 in total

1.  The history of extracorporeal oxygenators.

Authors:  M W Lim
Journal:  Anaesthesia       Date:  2006-10       Impact factor: 6.955

2.  Sedative clearance during extracorporeal membrane oxygenation.

Authors:  Varsha Bhatt-Meht; Gail Annich
Journal:  Perfusion       Date:  2005-10       Impact factor: 1.972

3.  Blood compatibility of methyl, methyl vinyl, methyl phenyl, and trifluoropropylmethylvinyl silicone rubber without silica fillers in the spiral-coiled membrane lung.

Authors:  T Kolobow; T A Tomlinson; J E Pierce
Journal:  J Biomed Mater Res       Date:  1977-07

4.  Hemolysis after administration of high-dose fentanyl.

Authors:  H Furuya; F Okumura
Journal:  Anesth Analg       Date:  1986-02       Impact factor: 5.108

5.  In vitro evaluation of sedative drug losses during extracorporeal membrane oxygenation.

Authors:  H Mulla; G Lawson; C von Anrep; M D Burke; D U Upton; R K Firmin; H Killer
Journal:  Perfusion       Date:  2000-01       Impact factor: 1.972

Review 6.  Pharmacokinetic changes during extracorporeal membrane oxygenation: implications for drug therapy of neonates.

Authors:  Marcia L Buck
Journal:  Clin Pharmacokinet       Date:  2003       Impact factor: 6.447

7.  Preliminary studies of the effects of extracorporeal membrane oxygenator on the disposition of common pediatric drugs.

Authors:  O Dagan; J Klein; C Gruenwald; D Bohn; G Barker; G Koren
Journal:  Ther Drug Monit       Date:  1993-08       Impact factor: 3.681

8.  In vivo uptake and elimination of isoflurane by different membrane oxygenators during cardiopulmonary bypass.

Authors:  Christoph Wiesenack; Gunther Wiesner; Cornelius Keyl; Michael Gruber; Alois Philipp; Markus Ritzka; Christopher Prasser; Kai Taeger
Journal:  Anesthesiology       Date:  2002-07       Impact factor: 7.892

9.  Fentanyl plasma levels after modified ultrafiltration in infant heart surgery.

Authors:  Andreas H Taenzer; Robert Groom; Reed D Quinn
Journal:  J Extra Corpor Technol       Date:  2005-12
  9 in total
  17 in total

Review 1.  Contemporary extracorporeal membrane oxygenation for adult respiratory failure: life support in the new era.

Authors:  Graeme MacLaren; Alain Combes; Robert H Bartlett
Journal:  Intensive Care Med       Date:  2011-12-07       Impact factor: 17.440

2.  Extracorporeal tubing in the roller pump raceway: physical changes and particulate generation.

Authors:  Allison J Bednarski Spiwak; Alexander Horbal; Robert Leatherbury; Derek J Hansford
Journal:  J Extra Corpor Technol       Date:  2008-09

Review 3.  Considerations for analgosedation and antithrombotic management during extracorporeal life support.

Authors:  Pamela K Burcham; Alan J Rozycki; Erik E Abel
Journal:  Ann Transl Med       Date:  2017-02

4.  Local Anesthetic Systemic Toxicity Induced Cardiac Arrest After Topicalization for Transesophageal Echocardiography and Subsequent Treatment With Extracorporeal Cardiopulmonary Resuscitation.

Authors:  Brandon Bacon; Natalie Silverton; Micah Katz; Elise Heath; David A Bull; Jason Harig; Joseph E Tonna
Journal:  J Cardiothorac Vasc Anesth       Date:  2018-01-31       Impact factor: 2.628

5.  Modified surface coatings and their effect on drug adsorption within the extracorporeal life support circuit.

Authors:  Thomas J Preston; Todd M Ratliff; Daniel Gomez; Vincent E Olshove; Kathleen K Nicol; Cheryl L Sargel; Louis G Chicoine
Journal:  J Extra Corpor Technol       Date:  2010-09

Review 6.  Pharmacokinetics and Dosing of Anti-infective Drugs in Patients on Extracorporeal Membrane Oxygenation: A Review of the Current Literature.

Authors:  Jennifer Sherwin; Travis Heath; Kevin Watt
Journal:  Clin Ther       Date:  2016-08-21       Impact factor: 3.393

7.  Amiodarone Extraction by the Extracorporeal Membrane Oxygenation Circuit.

Authors:  C Griffin McDaniel; C Cole Honeycutt; Kevin M Watt
Journal:  J Extra Corpor Technol       Date:  2021-03

8.  Intravenous Ribavirin for Parainfluenza and Respiratory Syncytial Virus in an Infant Receiving Extracorporeal Membrane Oxygenation and Continuous Renal Replacement Therapy.

Authors:  Cheryl L Sargel; Mohammed Aboud; Aimee Forster; Loralie J Langman; Jessica Tansmore; Bruce A Mueller; William E Smoyer; Jeffery J Auletta
Journal:  J Pediatr Pharmacol Ther       Date:  2018 Jul-Aug

Review 9.  Clinical pharmacology considerations for children supported with ventricular assist devices.

Authors:  Jennifer Sherwin; Elizabeth Thompson; Kevin D Hill; Kevin Watt; Andrew J Lodge; Daniel Gonzalez; Christoph P Hornik
Journal:  Cardiol Young       Date:  2018-07-11       Impact factor: 1.093

10.  Antifungal Extraction by the Extracorporeal Membrane Oxygenation Circuit.

Authors:  Kevin M Watt; Michael Cohen-Wolkowiez; Duane C Williams; Desiree K Bonadonna; Ira M Cheifetz; Dhiren Thakker; Daniel K Benjamin; Kim L R Brouwer
Journal:  J Extra Corpor Technol       Date:  2017-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.