Literature DB >> 18291398

A novel biodegradable pancreatic stent for human pancreatic applications: a preclinical safety study in a large animal model.

Johanna Laukkarinen1, Teemu Lämsä, Isto Nordback, Joonas Mikkonen, Juhani Sand.   

Abstract

BACKGROUND: Endoscopic stenting is one treatment method for pancreatic strictures or pseudocysts in patients with symptomatic chronic pancreatitis. With a biodegradable stent, the later removal of the stent could be avoided.
OBJECTIVE: We investigated the degradation, patency, and toxicity of a novel biodegradable, self-expanding radiopaque polylactide-barium sulfate pancreatic stent in a large animal model.
DESIGN: Animal study. SETTING AND
INTERVENTIONS: Five swine had a biodegradable pancreatic stent placed into the pancreatic duct (PD) and were followed-up for 6 months. MAIN OUTCOME MEASUREMENTS: Repeated blood tests and radiographs were studied during the follow-up. The animals were euthanized at 6 months, at which time, the PD inner diameter was measured, and histology was analyzed and graded. For comparison, histology from 5 nonstented animals was also analyzed.
RESULTS: The stent was correctly inserted into the PD in 4 of 5 animals. All the animals remained healthy, gained weight, and showed no signs of pancreatitis. A radiograph showed that the stent was in its original place in all animals at 1 month but had disappeared in all animals by 3 months. At 6 months, the autopsied pancreatic tissue was soft, and the PD was patent in all of the animals. The PD was slightly dilated at the site of the stent in the head of the pancreas compared with the preoperative diameter (5 mm [range 3-6 mm] vs 2 mm [range 1-3 mm], P < .05), but, in the body of the pancreas, no significant dilatation was seen. In the histology of the pancreata, there was no difference between the samples from the exposed parts of the biodegradable pancreatic stent compared with the distal nonexposed parts, or to the samples from the nonstented animals.
CONCLUSIONS: This novel biodegradable pancreatic stent, studied in these swine, appeared to be safe for use in the PD. These encouraging results warrant further clinical trials with this biodegradable pancreatic stent in pancreatic applications in human beings.

Entities:  

Mesh:

Year:  2008        PMID: 18291398     DOI: 10.1016/j.gie.2007.10.013

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  8 in total

Review 1.  Biodegradable stents in gastrointestinal endoscopy.

Authors:  Vicente Lorenzo-Zúñiga; Vicente Moreno-de-Vega; Ingrid Marín; Jaume Boix
Journal:  World J Gastroenterol       Date:  2014-03-07       Impact factor: 5.742

2.  Biodegradable polydioxanone stents: a new option for therapy-resistant anastomotic strictures of the colon.

Authors:  Václav Janík; Ladislav Horák; Jan Hnaníček; Jiří Málek; Hans-Ulrich Laasch
Journal:  Eur Radiol       Date:  2011-04-30       Impact factor: 5.315

3.  A retrospective analysis of early and late outcome of biodegradable stent placement in the management of refractory anastomotic colorectal strictures.

Authors:  A Repici; N Pagano; G Rando; A Carlino; E Vitetta; E Ferrara; G Strangio; A Zullo; C Hassan
Journal:  Surg Endosc       Date:  2013-02-27       Impact factor: 4.584

Review 4.  Endoscopic therapy in chronic pancreatitis: current perspectives.

Authors:  Andrada Seicean; Simona Vultur
Journal:  Clin Exp Gastroenterol       Date:  2014-12-17

5.  BIODEGRADABLE BILIARY STENTS: A NEW APPROACH FOR THE MANAGEMENT OF HEPATICOJEJUNOSTOMY STRICTURES FOLLOWING BILE DUCT INJURY. PROSPECTIVE STUDY.

Authors:  Mariano E Giménez; Mariano Palermo; Eduardo Houghton; Pablo Acquafresca; Caetano Finger; Juan M Verde; Jorge Cardoso Cúneo
Journal:  Arq Bras Cir Dig       Date:  2016 Apr-Jun

6.  Radiopaque Chitosan Ducts Fabricated by Extrusion-Based 3D Printing to Promote Healing After Pancreaticoenterostomy.

Authors:  Maoen Pan; Chaoqian Zhao; Zeya Xu; Yuanyuan Yang; Tianhong Teng; Jinxin Lin; Heguang Huang
Journal:  Front Bioeng Biotechnol       Date:  2021-06-04

7.  Crystallization Behavior and Mechanical Properties of Poly(ε-caprolactone) Reinforced with Barium Sulfate Submicron Particles.

Authors:  Hegoi Amestoy; Paul Diego; Emilio Meaurio; Jone Muñoz; Jose-Ramon Sarasua
Journal:  Materials (Basel)       Date:  2021-05-02       Impact factor: 3.623

Review 8.  Futuristic Developments and Applications in Endoluminal Stenting.

Authors:  Joel Ferreira-Silva; Renato Medas; Mohit Girotra; Monique Barakat; James H Tabibian; Eduardo Rodrigues-Pinto
Journal:  Gastroenterol Res Pract       Date:  2022-01-11       Impact factor: 2.260

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.