A biomechanical and histological analysis of standard versus hydroxyapatite-coated pins for external fixation.

Antonio Moroni1, Matteo Cadossi, Matteo Romagnoli, Cesare Faldini, Sandro Giannini.   

Abstract

This sheep study was designed to make a comparative evaluation of two external fixation pin types each with and without hydroxyapatite (HA) coating. The two pins had different taper, pitch, and self drilling capabilities. Twenty Orthofix standard, self-tapping pins (group A), 20 Orthofix HA-coated, self-tapping pins (group B), 20 X-caliber, self-drilling, self-tapping pins (group C), and 20 X-caliber HA-coated, self-drilling, self-tapping pins (group D) were selected. Four pins were implanted in the right femurs of 20 adult sheep that were euthanized at 6 weeks. Mean pin insertion torque was 2745 +/- 822 Nmm in group A, 2726 +/- 784 Nmm in group B, 2818 +/- 552 Nmm in group C, and 2657 +/- 732 Nmm in group D (ns). Mean pin extraction torque was 1567 +/- 541 Nmm in group A, 2524 +/- 838 Nmm in group B, 1650 +/- 650 Nmm in group C, and 2517 +/- 726 Nmm in group D. HA-coated pins (group B and D) had a significantly greater mean pin extraction torque compared to similar uncoated pins (group A and C) (p < 0.0005). Histological analysis showed good osteointegration of the two coated pin types. This study shows that HA-coating is more important for optimal pin fixation than the particular combination of design parameters used in each pin type. (c) 2007 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18286593     DOI: 10.1002/jbm.b.31036

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  5 in total

Review 1.  Pedicle screw insertion techniques: an update and review of the literature.

Authors:  F Perna; R Borghi; F Pilla; N Stefanini; A Mazzotti; M Chehrassan
Journal:  Musculoskelet Surg       Date:  2016-11-19

2.  Titanium Kirschner Wires Resist Biofilms Better Than Stainless Steel and Hydroxyapatite-coated Wires: An In Vitro Study.

Authors:  James P McEvoy; Philip Martin; Arshad Khaleel; Shobana Dissanayeke
Journal:  Strategies Trauma Limb Reconstr       Date:  2019 May-Aug

3.  Improved bonding of partially osteomyelitic bone to titanium pins owing to biomimetic coating of apatite.

Authors:  Hirotaka Mutsuzaki; Yu Sogo; Ayako Oyane; Atsuo Ito
Journal:  Int J Mol Sci       Date:  2013-12-13       Impact factor: 5.923

Review 4.  From prevention of pin-tract infection to treatment of osteomyelitis during paediatric external fixation.

Authors:  Dimitri Ceroni; Catherine Grumetz; Odile Desvachez; Sophie Pusateri; Pierre Dunand; Eleftheria Samara
Journal:  J Child Orthop       Date:  2016-11-15       Impact factor: 1.548

5.  The effect of HIV infection on the incidence and severity of circular external fixator pin track sepsis: a retrospective comparative study of 229 patients.

Authors:  Nando Ferreira; Leonard Charles Marais
Journal:  Strategies Trauma Limb Reconstr       Date:  2014-07-24