Validation of a new noninvasive device for the monitoring of peak endocardial acceleration in pigs: implications for optimization of pacing site and configuration.

Pierre Bordachar1, Louis Labrousse, Sylvain Ploux, Jean-Benoit Thambo, Stephane Lafitte, Patricia Reant, Pierre Jais, Michel Haissaguerre, Jacques Clementy, Pierre Dos Santos.   

Abstract

INTRODUCTION: The peak of endocardial acceleration (PEA) is an index of myocardial contractility. We aimed to (1) demonstrate that the PEA measured by the noninvasive cutaneous precordial application of an accelerometer sensor is related to left ventricular (LV) dP/dt max and (2) assess the usefulness of PEA monitoring during graded ischemia and during different configurations of sequential biventricular pacing. METHODS AND
RESULTS: Measurements of invasive LV dP/dt max were compared with measurements of transcutaneous PEA in seven pigs at baseline and during acute drug infusions; increased heart rate; right, left, biventricular and sequential biventricular pacing before and after graded ischemia induced by the constriction of the left anterior descending coronary artery. A consistent PEA signal was obtained in all animals. PEA changes were highly related to LV dP/dt max changes (r= 0.93; P < 0.001). The changes of LV contractility induced by the different pacing configurations were detected by PEA analysis in the absence of ischemia (r= 0.94; P < 0.001) and in the presence of ischemic LV dysfunction (r= 0.91; P < 0.001).
CONCLUSION: Noninvasive PEA measurement allows monitoring of left ventricular contractility and may be a useful tool to detect global effect of ventricular ischemia and to optimize the choice of both pacing site and pacing configuration.

Entities:  

Mesh:

Year:  2008        PMID: 18284498     DOI: 10.1111/j.1540-8167.2008.01105.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  2 in total

1.  Left Ventricular Pressure Estimation Using Machine Learning-Based Heart Sound Classification.

Authors:  Philip Westphal; Hongxing Luo; Mehrdad Shahmohammadi; Luuk I B Heckman; Marion Kuiper; Frits W Prinzen; Tammo Delhaas; Richard N Cornelussen
Journal:  Front Cardiovasc Med       Date:  2022-05-25

2.  Heart failure monitoring with a cardiac resynchronization therapy device-based cardiac contractility sensor: a case series.

Authors:  Jacques Mansourati; Mélanie Heurteau; Jérôme Abaléa
Journal:  J Med Case Rep       Date:  2014-01-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.