BACKGROUND: Surgical navigation requires registration of the pre-operative image dataset with the patient in the operation theatre. Various marker and marker-free registration techniques are available, each bearing an individual level of precision and clinical practicability. In this study the precision of four different registration methods in a maxillofacial surgical setting is analyzed.
MATERIALS AND METHODS: A synthetic full size human skull model was registered with its computer tomography-dataset using (a) a dentally mounted occlusal splint, (b) the laser surface scanning, (c) five facial bone implants and (d) a combination of dental splint and two orbital bone implants. The target registration error was computed for 170 landmarks spread over the entire viscero- and neurocranium in 10 repeats using the VectorVision2 (BrainLAB AG, Heimstetten, Germany) navigation system. Statistical and graphical analyses were performed by anatomical region.
RESULTS: An average precision of 1mm was found for the periorbital region irrespective of registration method (range 0.6-1.1mm). Beyond the mid-face, precision linearly decreases with the distance from the reference markers. The combination of splint with two orbital bone markers significantly improved precision from 1.3 to 0.8mm (p<0.001) on the viscerocranium and 2.3-1.2mm (p<0.001) on the neurocranium.
CONCLUSIONS: An occlusal splint alone yields poor precision for navigation beyond the mid-face. The precision can be increased by combining an occlusal splint with just two bone implants inserted percutaneously on the lateral orbital rim of each side.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18280173     DOI: 10.1016/j.jcms.2007.09.002

Source DB:  PubMed          Journal:  J Craniomaxillofac Surg        ISSN: 1010-5182            Impact factor:   2.078


  38 in total

1.  Marker-free registration for the accurate integration of CT images and the subject's anatomy during navigation surgery of the maxillary sinus.

Authors:  S-H Kang; M-K Kim; J-H Kim; H-K Park; W Park
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

2.  Titanium preformed implants in orbital floor reconstruction - case presentation, review of literature.

Authors:  Bogdan Banica; Patricia Ene; Daniela Vranceanu; Razvan Ene
Journal:  Maedica (Bucur)       Date:  2013-03

3.  Is there an equivalence of non-invasive to invasive referenciation in computer-aided surgery?

Authors:  Tanja D Grauvogel; Juergen Grauvogel; Susan Arndt; Ansgar Berlis; Wolfgang Maier
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-05-05       Impact factor: 2.503

4.  An effective point-based registration tool for surgical navigation.

Authors:  Jaesung Hong; Makoto Hashizume
Journal:  Surg Endosc       Date:  2009-06-16       Impact factor: 4.584

5.  Warning navigation system using real-time safe region monitoring for otologic surgery.

Authors:  Byunghyun Cho; Masamichi Oka; Nozomu Matsumoto; Riichi Ouchida; Jaesung Hong; Makoto Hashizume
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-11-16       Impact factor: 2.924

6.  Application of a real-time three-dimensional navigation system to various oral and maxillofacial surgical procedures.

Authors:  Seigo Ohba; Hitoshi Yoshimura; Kyoko Ishimaru; Kousuke Awara; Kazuo Sano
Journal:  Odontology       Date:  2014-05-09       Impact factor: 2.634

7.  Registration using 3D-printed rigid templates outperforms manually scanned surface matching in image-guided temporal bone surgery.

Authors:  Makoto Yamashita; Nozomu Matsumoto; Byunghyun Cho; Noritaka Komune; Shinya Onogi; Jongseung Lee; Jordan Bano; Tomohiko Akahoshi; Makoto Hashizume
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-06-14       Impact factor: 2.924

8.  Prospects and limitations of different registration modalities in electromagnetic ENT navigation.

Authors:  Eric Soteriou; Juergen Grauvogel; Roland Laszig; Tanja Daniela Grauvogel
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-05-05       Impact factor: 2.503

9.  Intraoperative navigation in complex head and neck resections: indications and limits.

Authors:  S Catanzaro; C Copelli; A Manfuso; K Tewfik; N Pederneschi; L Cassano; R Cocchi
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-09-22       Impact factor: 2.924

10.  Performance of Robotic Assistance for Skull Base Biopsy: A Phantom Study.

Authors:  Jian-Hua Zhu; Jing Wang; Yong-Gui Wang; Meng Li; Yu-Xing Guo; Xiao-Jing Liu; Chuan-Bin Guo
Journal:  J Neurol Surg B Skull Base       Date:  2017-05-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.