Literature DB >> 18278730

Multipolar hepatic radiofrequency ablation using up to six applicators: preliminary results.

P Bruners1, T Schmitz-Rode, R W Günther, A Mahnken.   

Abstract

PURPOSE: To evaluate the clinical feasibility and safety of hepatic radiofrequency (RF) ablation using a multipolar RF system permitting the simultaneous use of up to six electrodes.
MATERIALS AND METHODS: Ten patients (3 female, 7 male, mean age 61) suffering from 29 hepatic metastases (range: 1 - 5) of different tumors were treated with a modified multipolar RF system (CelonLab Power, Celon Medical Instruments, Teltow, Germany) operating four to six needle-shaped internally cooled RF applicators. The procedure duration, applied energy and generator output were recorded during the intervention. The treatment result and procedure-related complications were analyzed. The achieved coagulation volume was calculated on the basis of contrast-enhanced CT scans 24 hours after RF ablation.
RESULTS: Complete tumor ablation was achieved in all cases as determined by the post-interventional lack of contrast enhancement in the target region using four applicators in five patients, five applicators in one patient and six applicators in four patients. A mean energy deposition of 353.9 +/- 176.2 kJ resulted in a mean coagulation volume of 115.9 +/- 79.5 cm (3). The mean procedure duration was 74.9 +/- 21.2 minutes. Four patients showed an intraabdominal hemorrhage which necessitated further interventional treatment (embolization; percutaneous histoacryl injection) in two patients.
CONCLUSION: Multipolar RF ablation of hepatic metastasis with up to six applicators was clinically feasible. In our patient population it was associated with an increased risk of intraabdominal bleeding probably due to the multiple punctures associated with the use of multiple applicators.

Entities:  

Mesh:

Year:  2008        PMID: 18278730     DOI: 10.1055/s-2008-1027184

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  4 in total

1.  Comparison of four radiofrequency ablation systems at two target volumes in an ex vivo bovine liver model.

Authors:  Hendrik Rathke; Bernd Hamm; Felix Güttler; Joern Rathke; Jens Rump; Ulf Teichgräber; Maximilian de Bucourt
Journal:  Diagn Interv Radiol       Date:  2014 May-Jun       Impact factor: 2.630

Review 2.  Tumour ablation: technical aspects.

Authors:  Gerlig Widmann; Gerd Bodner; Reto Bale
Journal:  Cancer Imaging       Date:  2009-10-02       Impact factor: 3.909

3.  Radiofrequency ablation versus hepatic resection for breast cancer liver metastasis: a systematic review and meta-analysis.

Authors:  Yi-Bin Xiao; Bo Zhang; Yu-Lian Wu
Journal:  J Zhejiang Univ Sci B       Date:  2018-11       Impact factor: 3.066

4.  Percutaneous Dual-Switching Monopolar Radiofrequency Ablation Using a Separable Clustered Electrode: A Preliminary Study.

Authors:  Tae Won Choi; Jeong Min Lee; Dong Ho Lee; Jeong-Hoon Lee; Su Jong Yu; Yoon Jun Kim; Jung-Hwan Yoon; Joon Koo Han
Journal:  Korean J Radiol       Date:  2017-07-17       Impact factor: 3.500

  4 in total