Literature DB >> 18267313

An improved alternative to vacuum-assisted closure (VAC) as a negative pressure dressing in lower limb split skin grafting: a clinical trial.

Warren Matthew Rozen1, Shekhib Shahbaz, Adel Morsi.   

Abstract

BACKGROUND: The use of negative pressure in the dressing of split skin grafts has been shown to promote healing by a variety of mechanisms, including a decrease in interstitial oedema, an increase in perfusion and a decrease in bacterial colonisation. Vacuum-assisted closure (VAC) dressings have, until now, been used as the archetype for negative pressure dressings and have been reflected as such in the literature. However, patient mobility and cost are still an issue with these dressings, and alternatives have been keenly sought. We describe an alternative method of negative pressure dressing, which we have found to be a safe and successful alternative in the setting of lower limb split skin grafts.
MATERIALS AND METHODS: A prospective cohort investigation was performed on nine consecutive patients at Monash Medical Centre undergoing split-skin grafting for a lower limb soft tissue defect. The dressing comprised a single cut foam sheet, a conventional disposable closed-system suction drain and an adhesive dressing.
RESULTS: In all nine patients, there was a 100% take of the graft, with no partial or complete loss. There were no complications encountered. Cost analysis demonstrated a minimum treatment cost of $577 over 5 days compared to $3180 for commercial VAC dressed wounds: a net saving of $2603 per patient.
CONCLUSIONS: The use of a simple suction drain is a cheap and safe alternative to commercial VAC dressings for the treatment of lower limb split skin grafts. Length of hospital stay and cost are superior to VAC, with no diminished clinical outcome.

Entities:  

Mesh:

Year:  2007        PMID: 18267313     DOI: 10.1016/j.bjps.2007.01.064

Source DB:  PubMed          Journal:  J Plast Reconstr Aesthet Surg        ISSN: 1748-6815            Impact factor:   2.740


  4 in total

1.  Vacuum-assisted closure improves the incorporation of artificial dermis in soft tissue defects: Terudermis(®) and Pelnac(®).

Authors:  SuRak Eo; YoongSoo Kim; SangHun Cho
Journal:  Int Wound J       Date:  2011-02-25       Impact factor: 3.315

2.  Role of customised negative-pressure wound therapy in the integration of split-thickness skin grafts: A randomised control study.

Authors:  Mir Mohsin; Haroon Rashid Zargar; Adil Hafeez Wani; Mohammad Inam Zaroo; Peerzada Umar Farooq Baba; Sheikh Adil Bashir; Altaf Rasool; Akram Hussain Bijli
Journal:  Indian J Plast Surg       Date:  2017 Jan-Apr

3.  Roles of Oxidative Stress and Raftlin in Wound Healing Under Negative-Pressure Wound Therapy.

Authors:  Xingan Qiu; Yifan Wu; Dong Zhang; Hao Zhang; Aixi Yu; Zonghuan Li
Journal:  Clin Cosmet Investig Dermatol       Date:  2021-11-20

4.  Application of the Single Use Negative Pressure Wound Therapy Device (PICO) on a Heterogeneous Group of Surgical and Traumatic Wounds.

Authors:  Caroline Payne; Daren Edwards
Journal:  Eplasty       Date:  2014-04-28
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.