Selections of the cylinder implant neck taper and implant end fillet for optimal biomechanical properties: a three-dimensional finite element analysis.

Liang Kong1, Yingying Sun, Kaijin Hu, Yanpu Liu, Dehua Li, Zhihui Qiu, Baolin Liu.   

Abstract

In this paper, effects of the implant neck taper and the implant-end fillet on the maximum Von Mises stresses were evaluated in jaw bones, and maximum displacements examined in an implant-abutment complex by a finite element method (FEM). The implant-neck tapers (T) ranged from 45 degrees to 70 degrees , and fillets of implant ends (R) ranged from 0.5 to 1.5mm. Results suggested that under axial load by the maximum Von Mises stresses in cortical and cancellous bones decreased by 71.6% and 14.8%, respectively, and under 45 degrees buccolingual load by 68.2% and 11.0%, respectively. The maximum displacement of implant-abutment complex decreased by 9.1% and 22.8% under axial and 45 degrees buccolingual load, respectively. When T ranged from 64 degrees to 73 degrees and R exceeded 0.8mm, minimum stress/displacement was obtained and the evaluating targets were more sensitive to T than to R. Data indicated that the taper of implant neck favored stress distribution in cortical bones more than the fillet of implant end did; taper of implant neck affected implant stability more than the fillet of implant end did; and the taper of implant neck of 64-73 degrees and fillet of implant end exceeding 0.8mm were optimal selections for the type B/2 bone in a cylinder implant by biomechanical consideration.

Mesh:

Substances:

Year:  2008        PMID: 18258239     DOI: 10.1016/j.jbiomech.2007.12.013

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  4 in total

1.  Biomechanical role of peri-implant trabecular structures during vertical loading.

Authors:  Takashi Ohashi; Satoru Matsunaga; Ken Nakahara; Shinichi Abe; Yoshinobu Ide; Yuichi Tamatsu; Naoki Takano
Journal:  Clin Oral Investig       Date:  2009-08-18       Impact factor: 3.573

2.  Effect of abutment neck taper and cement types on the amount of remnant cement in cement-retained implant restorations: an in vitro study.

Authors:  Yeon-Hee Park; Kyoung-A Kim; Jung-Jin Lee; Tae-Min Kwon; Jae-Min Seo
Journal:  J Adv Prosthodont       Date:  2022-06-27       Impact factor: 1.989

3.  Relationships of Stresses on Alveolar Bone and Abutment of Dental Implant from Various Bite Forces by Three-Dimensional Finite Element Analysis.

Authors:  Xiaoning Kang; Yiming Li; Yixi Wang; Yao Zhang; Dongsheng Yu; Yun Peng
Journal:  Biomed Res Int       Date:  2020-02-19       Impact factor: 3.411

4.  Development of Response Surface Model of Endurance Time and Structural Parameter Optimization for a Tailsitter UAV.

Authors:  Xiaomin Yao; Wenshuai Liu; Wenting Han; Guang Li; Qian Ma
Journal:  Sensors (Basel)       Date:  2020-03-22       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.