Validation study of nasal nitric oxide measurements using a hand-held electrochemical analyser.

M Maniscalco1, G de Laurentiis, E Weitzberg, J O Lundberg, M Sofia.   

Abstract

BACKGROUND: Exhaled nitric oxide (NO) measurement is a simple and non-invasive method for monitoring airway inflammation. Similarly, nasal NO has been proposed as a surrogate marker in inflammatory diseases of the upper airways, e.g. allergic rhinitis. A new portable analyser using an electrochemical sensor has been developed for measurements of exhaled NO, and its reproducibility and comparison with other analysers has been tested recently in healthy subjects and in patients with lower airways disease. The application of this hand-held analyser in nasal NO analysis was tested and compared to the gold standard represented by a chemiluminescence analyser.
MATERIALS AND METHODS: Thirty subjects including 15 patients with allergic rhinitis (AR) and 15 healthy subjects (HS) were studied. The intraindividual variability, calculated as the difference in nasal NO levels between two measurements from a single nasally exhaled breath manoeuvre, and the comparison between the electrochemical analyser (NIOX MINO, Aerocrine) and a chemiluminescence analyser (NOA, Sievers) were performed.
RESULTS: In AR patients mean nasal NO was 59.0 +/- 16.3 p.p.b. with the MINO and 58.3 +/- 15.6 p.p.b. with the NOA. In HS nasal NO was 49.1 +/- 10.8 p.p.b. with the MINO and 49.8 +/- 8.2 p.p.b. with the NOA. The Bland-Altman analysis showed bias values of 0.005 +/- 3.6 with the 95% limits of agreement from -6.97 to 6.98 p.p.b.
CONCLUSION: Measurements of nasal NO levels with a hand-held electrochemical analyser are reproducible and the results are comparable to a stationary chemiluminescence analyser.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18257783     DOI: 10.1111/j.1365-2362.2008.01907.x

Source DB:  PubMed          Journal:  Eur J Clin Invest        ISSN: 0014-2972            Impact factor:   4.686


  8 in total

Review 1.  An Evidence-Based Review of Application Devices for Nitric Oxide Concentration Determination from Exhaled Air in the Diagnosis of Inflammation and Treatment Monitoring.

Authors:  Magdalena Wyszyńska; Monika Nitsze-Wierzba; Aleksandra Czelakowska; Jacek Kasperski; Joanna Żywiec; Małgorzata Skucha-Nowak
Journal:  Molecules       Date:  2022-07-03       Impact factor: 4.927

2.  Hand-held tidal breathing nasal nitric oxide measurement--a promising targeted case-finding tool for the diagnosis of primary ciliary dyskinesia.

Authors:  June Kehlet Marthin; Kim Gjerum Nielsen
Journal:  PLoS One       Date:  2013-02-20       Impact factor: 3.240

3.  Validation of a portable nitric oxide analyzer for screening in primary ciliary dyskinesias.

Authors:  Amanda Harris; Esther Bhullar; Kerry Gove; Rhiannon Joslin; Jennifer Pelling; Hazel J Evans; Woolf T Walker; Jane S Lucas
Journal:  BMC Pulm Med       Date:  2014-02-10       Impact factor: 3.317

4.  Clinical Utility Of The Exhaled Nitric Oxide (NO) Measurement With Portable Devices In The Management Of Allergic Airway Inflammation And Asthma.

Authors:  Sy Duong-Quy
Journal:  J Asthma Allergy       Date:  2019-10-07

5.  Pilot Study of Use of Nitric Oxide in Monitoring Multiple Dental Foci in Oral Cavity-A Case Report.

Authors:  Magdalena Wyszyńska; Przemysław Rosak; Aleksandra Czelakowska; Ewa Białożyt-Bujak; Jacek Kasperski; Maciej Łopaciński; Nour Al Khatib; Małgorzata Skucha-Nowak
Journal:  Healthcare (Basel)       Date:  2022-01-20

6.  Diagnostic accuracy of nasal nitric oxide for establishing diagnosis of primary ciliary dyskinesia: a meta-analysis.

Authors:  Panayiotis Kouis; Stefania I Papatheodorou; Panayiotis K Yiallouros
Journal:  BMC Pulm Med       Date:  2015-12-03       Impact factor: 3.317

7.  Clinical Distinctness of Allergic Rhinitis in Patients with Allergy to Molds.

Authors:  Krzysztof Kołodziejczyk; Andrzej Bozek
Journal:  Biomed Res Int       Date:  2016-05-31       Impact factor: 3.411

Review 8.  Fractional exhaled nitric oxide-measuring devices: technology update.

Authors:  Mauro Maniscalco; Carolina Vitale; Alessandro Vatrella; Antonio Molino; Andrea Bianco; Gennaro Mazzarella
Journal:  Med Devices (Auckl)       Date:  2016-06-23
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.