Paraproteins are a common cause of interferences with automated chemistry methods.

Yusong Yang1, Peter J Howanitz, Joan H Howanitz, Helen Gorfajn, Keung Wong.   

Abstract

CONTEXT: Previous studies have shown that paraproteins caused spurious results on individual analytes including total bilirubin (TBIL), direct bilirubin (DBIL), or HDL-cholesterol (HDL-C). Studies demonstrating paraprotein interferences with multiple analytes measured by different analyzers have not been reported.
OBJECTIVE: To systemically investigate interferences of paraproteins on TBIL, DBIL, and HDL-C measured by the Roche MODULAR and the Olympus AU2700.
DESIGN: Eighty-eight serum specimens with monoclonal gammopathies were analyzed using the Roche MODULAR and the Olympus AU2700. Paraprotein interferences with the MODULAR and AU2700 were identified by abnormal absorbance curves and confirmed by results from the Ortho Vitros 950 or inconsistent laboratory information.
RESULTS: Spurious results occurred in 89 of 528 measurements; 29 specimens did not demonstrate any interferences whereas 26 specimens gave spurious results in 2 to 4 of the 6 assays. Paraprotein interferences caused spuriously high levels of TBIL in 4 sera measured by the MODULAR. In contrast, paraprotein interferences on DBIL were observed by at least 1 method in 44% (39/88) of sera assayed, occurring almost exclusively with the AU2700. Paraprotein interferences with HDL-C results were present in 35% of specimens assayed with the MODULAR and 16% of specimens assayed with the AU2700. In specimens with interferences, spuriously low AU2700 DBIL, MODULAR HDL-C, and AU2700 HDL-C results occurred with 28%, 90%, and 91% of specimens, respectively.
CONCLUSIONS: We demonstrated that paraprotein interferences with TBIL, DBIL, and HDL-C are relatively common and provided explanations why these interferences occurred. Although it is difficult to predict which specimens cause interferences, spurious results appeared method and concentration dependent.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18251580     DOI: 10.5858/2008-132-217-PAACCO

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  9 in total

1.  Spurious hyperphosphatemia in a case of multiple myeloma.

Authors:  Sutirtha Chakraborty; Susruta Sen; Debkishore Gupta; Sidhartha Sankar Ghosh; Prasad Sawant; Mandrita Das
Journal:  Indian J Clin Biochem       Date:  2013-08-08

2.  A curious case of hyperbilirubinemia.

Authors:  Atanu Kumar Dutta
Journal:  Indian J Clin Biochem       Date:  2011-09-21

3.  A component could not be larger than the whole, polyclonal immunoglobulin interference in conjugated/direct bilirubin assay and elimination: A case report.

Authors:  Neda Soleimani; Saeideh Khaleghpanah; Sahand Mohammadzadeh
Journal:  Clin Case Rep       Date:  2022-05-15

4.  Unusual manifestation of monoclonal gammopathy of undetermined significance: a false serum creatinine elevation.

Authors:  Yoshinosuke Shimamura; Takuto Maeda; Yayoi Ogawa; Yuki Nagai; Toshiya Shinohara; Hideki Takizawa
Journal:  CEN Case Rep       Date:  2019-12-14

5.  Artifactual measurement of low serum HDL-cholesterol due to paraproteinemia.

Authors:  Edwin O F van Gorselen; Theo Diekman; Jan Hessels; Patrick M J Verhorst; Clemens von Birgelen
Journal:  Clin Res Cardiol       Date:  2010-07-01       Impact factor: 5.460

6.  Paraprotein interference with turbidimetric gentamicin assay.

Authors:  Goce Dimeski; Kendra Bassett; Nigel Brown
Journal:  Biochem Med (Zagreb)       Date:  2015       Impact factor: 2.313

7.  Deceiving proteins! A case of lymphoma and high creatinine.

Authors:  El Hakem Abdelkarim Metraiah; Helen Regan; Johanna Louw; Dana Kidder
Journal:  BMJ Case Rep       Date:  2017-01-23

8.  The pH of chemistry assays plays an important role in monoclonal immunoglobulin interferences.

Authors:  Michael O Alberti; Thomas A Drake; Lu Song
Journal:  Pract Lab Med       Date:  2015-10-09

9.  Operational impact of using a vanadate oxidase method for direct bilirubin measurements at an academic medical center clinical laboratory.

Authors:  Neha Dhungana; Cory Morris; Matthew D Krasowski
Journal:  Pract Lab Med       Date:  2017-05-17
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.