| Literature DB >> 18243979 |
G T Gullberg, C R Crawford, B M Tsui.
Abstract
A convolutional backprojection algorithm is derived for a fan beam geometry that has its center-of-rotation displaced from the midline of the fan beam. In single photon emission computed tomography (SPECT), where a transaxial converging collimator is used with a rotating gamma camera, it is difficult to precisely align the collimator so that the mechanical center-of-rotation is colinear with the midline of the fan beam. A displacement of the center-of-rotation can also occur in X-ray CT when the X-ray source is mispositioned. Standard reconstruction algorithms which directly filter and backproject the fan beam data without rebinning into parallel beam geometry have been derived for a geometry having its center-of-rotation at the midline of the fan beam. However, in the case of a misalignment of the center-of-rotation, if these conventional reconstruction algorithms are used to reconstruct the fan beam projections, structured artifacts and a loss of resolution will result. We illustrate these artifacts with simulations and demonstrate how the news algorithm corrects for this misalignment. We also show a method to estimate the parameters of the fan beam geometry including the shift in the center-of-rotation.Entities:
Year: 1986 PMID: 18243979 DOI: 10.1109/TMI.1986.4307735
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048