| Literature DB >> 18238864 |
Bart E Muhs1, Koen L Vincken, Arno Teutelink, Eric L G Verhoeven, Matthias Prokop, Frans L Moll, Hence J M Verhagen.
Abstract
Different endograft configurations (fenestrated, transrenal, infrarenal) may varyingly affect aortic side branch movement. Renal artery motion was evaluated with 64-slice dynamic cine-computed tomography angiography before and after endovascular aneurysm repair in 16 patients (46 renal arteries). Center-of-mass displacement of the renals was determined per heartbeat for before repair for 3 different endografts; differences were compared, with significance at P < 0.5. Preoperative renal artery motion is significant (1.2 [SD 0.5] mm, range, 0.6-2.). Neither transrenal nor infrarenal endografts alter renal artery motion compared with before repair (P < .05). Renal artery motion after fenestrated endovascular repair with renal stents reduces motion to 25% of the preoperative value (0.3 [SD, 0.1] mm, range, 0.2-0.5 mm; P = .01). Endograft implantation without stented side branches does not change renal artery motion, potentially allowing significant movement of the renal artery relative to the fenestration. Routine stenting of fenestrations limits postoperative renal artery motion to 0.3 mm, thereby preventing significant branch movement in relation the fenestration.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18238864 DOI: 10.1177/1538574407308200