Relationship between implant stability measured by resonance frequency analysis (RFA) and bone loss during early healing period.

Seung-Min Yang1, Seung-Yun Shin, Seung-Beom Kye.   

Abstract

OBJECTIVE: The aim of this study was to monitor the changes in the stability-related bone loss during the early healing period. STUDY
DESIGN: A total of 43 implants were installed in 19 patients, and their stability was measured by resonance frequency analysis (RFA) at the time of surgery and once per week for 12 weeks. A total of 32 implants had radiographs taken at the placement of implant and 4, 8, and 12 weeks postoperatively. The data for gender, bone density, different surgery protocols, and bone loss around implant were analyzed.
RESULTS: The mean implant stability quotient values were not changed during the healing period. The marginal bone losses at 12 weeks were 1.28 +/- 0.51 mm and 1.32 +/- 0.57 mm on the mesial and distal sides, respectively.
CONCLUSIONS: The implant stability in normal and hard bone density did not change with time, and there was no correlation between the marginal bone loss and the change of implant stability.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18230372     DOI: 10.1016/j.tripleo.2007.08.045

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  6 in total

1.  The effect of peri-implant bone exposure on soft tissue healing and bone loss in two adjacent implants.

Authors:  Seung-Yun Shin; Seung-Boem Kye; Jongrak Hong; Jun-Young Paeng; Seung-Min Yang
Journal:  J Periodontal Implant Sci       Date:  2012-02-29       Impact factor: 2.614

Review 2.  Clinical evaluation of osseointegration using resonance frequency analysis.

Authors:  Parth Satwalekar; Sandeep Nalla; Ramaswamy Reddy; Sheeba Glory Chowdary
Journal:  J Indian Prosthodont Soc       Date:  2015 Jul-Sep

3.  Relation between the stability of dental implants and two biological markers during the healing period: a prospective clinical study.

Authors:  Choknapa Tirachaimongkol; Peraphan Pothacharoen; Peter A Reichart; Pathawee Khongkhunthian
Journal:  Int J Implant Dent       Date:  2016-12-08

4.  Damping ratio analysis of tooth stability under various simulated degrees of vertical alveolar bone loss and different root types.

Authors:  Kuo-Ning Ho; Sheng-Yang Lee; Haw-Ming Huang
Journal:  Biomed Eng Online       Date:  2017-08-03       Impact factor: 2.819

5.  Reliability of the Resonance Frequency Analysis Values in New Prototype Transepithelial Abutments: A Prospective Clinical Study.

Authors:  María Guerrero-González; Francesca Monticelli; David Saura García-Martín; Mariano Herrero-Climent; Blanca Ríos-Carrasco; José-Vicente Ríos-Santos; Ana Fernández-Palacín
Journal:  Int J Environ Res Public Health       Date:  2020-09-16       Impact factor: 3.390

6.  Evaluation of Implant Stability and Trephination Depth for Implant Removal-An In Vitro Study.

Authors:  Haya Meir; Alon Sebaoun; Perry Raz; Shifra Levartovsky; Adi Arieli; Raphel Pilo; Zafar Dor; Ilan Beitlitum
Journal:  Materials (Basel)       Date:  2022-06-13       Impact factor: 3.748

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.