The authors present the count rate performance of the CEA-LETI TTVO3 time-of-flight positron emission tomography (PET) system using both physical measurements and H(2)(15)O bolus human myocardial studies. They also present a comparison between the counting statistics of H(2)(15)O brain studies performed on this machine and on the latest available high-resolution brain bismuth germanate (BGO) tomograph, the ECAT 953B/31. During the 80 mCi cerebral blood flow study, the count rate reached 100 K events/s, and the same experiment performed on a high-resolution BGO brain machine gave only a 30% increase in signal. These results demonstrate that TTVO3 is particularly suitable for H(2)(15)O flow studies.

Entities:  

Year:  1991        PMID: 18222826     DOI: 10.1109/42.97574

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  1 in total

1.  Impairment of the exercise-induced increase in muscle perfusion in McArdle's disease.

Authors:  P Jehenson; A Leroy-Willig; E de Kerviler; P Merlet; D Duboc; A Syrota