T F Foo1, N S Grigsby, J D Mitchell, B E Slayman.
Abstract
A method for detection and removal of random spike noise in magnetic resonance (MR) raw data (k-space data) is described. This method would reduce or eliminate the corduroy-type and higher than usual level artifacts in MR images resulting from random spike noise in k-space data. The method described involves applying a spatially varying threshold to be k-space data. Any data point that has a magnitude greater than that of the threshold value at that location will be replaced by a local complex average of the neighboring data points or some other suitable data replacement scheme.Year: 1994 PMID: 18218490 DOI: 10.1109/42.276151
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048