A new experimental method for measuring skin's natural tension.

Emmanuelle Jacquet1, Gwendal Josse, Fouad Khatyr, Camille Garcin.   

Abstract

PURPOSE: The precise determination of skin's mechanical properties is still an open question. When performing an in vivo test, the piece of skin tested is not as well defined as it is in material testing. Moreover, the body zone and the body posture imply an initial stress on the skin. Consequently, a precise mechanical analysis needs a precise measurement of the natural skin tension.
METHODS: A new method and the relative device are presented. It is based on an extensiometry test. Skin is tested not only in traction but also in compression. The tested skin sample is well defined and protected from surrounding effects by follower tabs. The size and shape of the device have been optimised by a finite element modelisation.
RESULTS: The method was tested with elastomers pre-tensioned at different loads. It is shown that the initial tension can be retrieved with good precision. Tests were then performed in vivo on the forearm for different arm positions. It is shown that initial tension could be only clearly determined for the highest skin tension, although the skin presented very different traction behaviour with different arm positions.
CONCLUSION: It is shown how body posture influences measurements. An innovative method for easily determining initial tension is presented. Nevertheless, further tests and device improvements are needed to apply this skin tension measurement for different body zones and body postures.

Mesh:

Substances:

Year:  2008        PMID: 18211595     DOI: 10.1111/j.1600-0846.2007.00259.x

Source DB:  PubMed          Journal:  Skin Res Technol        ISSN: 0909-752X            Impact factor:   2.365


  5 in total

1.  The analysis of forces needed for the suturing of elliptical skin wounds.

Authors:  Lukas Capek; Emmanuelle Jacquet; Ladislav Dzan; Antonin Simunek
Journal:  Med Biol Eng Comput       Date:  2011-12-27       Impact factor: 2.602

2.  Miniaturized electromechanical devices for the characterization of the biomechanics of deep tissue.

Authors:  Enming Song; Zhaoqian Xie; Wubin Bai; Haiwen Luan; Bowen Ji; Xin Ning; Yu Xia; Janice Mihyun Baek; Yujin Lee; Raudel Avila; Huang-Yu Chen; Jae-Hwan Kim; Surabhi Madhvapathy; Kuanming Yao; Dengfeng Li; Jingkun Zhou; Mengdi Han; Sang Min Won; Xinyuan Zhang; Daniel J Myers; Yongfeng Mei; Xu Guo; Shuai Xu; Jan-Kai Chang; Xinge Yu; Yonggang Huang; John A Rogers
Journal:  Nat Biomed Eng       Date:  2021-05-27       Impact factor: 25.671

3.  The Compressiometer: Toward a New Skin Tensiometer for Research and Surgical Planning.

Authors:  Karlijn M J Scheepens; Nick Marsidi; Roel E Genders; Tim Horeman-Franse
Journal:  IEEE J Transl Eng Health Med       Date:  2021-12-06       Impact factor: 3.316

4.  Eyelid and Sternum Fibroblasts Differ in Their Contraction Potential and Responses to Inflammatory Cytokines.

Authors:  He Li; Jonathan C P Roos; Geoffrey E Rose; Maryse Bailly; Daniel G Ezra
Journal:  Plast Reconstr Surg Glob Open       Date:  2015-08-10

5.  A New Skin Tensiometer Device: Computational Analyses To Understand Biodynamic Excisional Skin Tension Lines.

Authors:  Sharad P Paul; Justin Matulich; Nick Charlton
Journal:  Sci Rep       Date:  2016-07-25       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.