Investigation of accelerated partial breast patient alignment and treatment with helical tomotherapy unit.

Katja M Langen1, Daniel J Buchholz, Doug R Burch, Rob Burkavage, Arti U Limaye, Sanford L Meeks, Patrick A Kupelian, Kenneth J Ruchala, Jason Haimerl, Doug Henderson, Gustavo H Olivera.   

Abstract

PURPOSE: To determine the precision of megavoltage computed tomography (MVCT)-based alignment of the seroma cavity for patients undergoing partial breast irradiation; and to determine whether accelerated partial breast irradiation (APBI) plans can be generated for TomoTherapy deliveries that meet the National Surgical Adjuvant Breast and Bowel Project (NSABP) B-39/Radiation Therapy Oncology Group (RTOG) 0413 protocol guidelines for target coverage and normal tissue dose limitations. METHODS AND MATERIALS: We obtained 50 MVCT images from 10 patients. An interuser study was designed to assess the alignment precision. Using a standard helical and a fixed beam prototype ("topotherapy") optimizer, two APBI plans for each patient were developed.
RESULTS: The precision of the MVCT-based seroma cavity alignment was better than 2 mm if averaged over the patient population. Both treatment techniques could be used to generate acceptable APBI plans for patients that fulfilled the recommended NSABP B-39/RTOG-0413 selection criteria. For plans of comparable treatment time, the conformation of the prescription dose to the target was greater for helical deliveries, while the ipsilateral lung dose was significantly reduced for the topotherapy plans.
CONCLUSIONS: The inherent volumetric imaging capabilities of a TomoTherapy Hi-Art unit allow for alignment of patients undergoing partial breast irradiation that is determined from the visibility of the seroma cavity on the MVCT image. The precision of the MVCT-based alignment was better than 2 mm (+/-standard deviation) when averaged over the patient population. Using the NSABP B-39/RTOG-0413 guidelines, acceptable APBI treatment plans can be generated using helical- or topotherapy-based delivery techniques.

Entities:  

Mesh:

Year:  2008        PMID: 18207666     DOI: 10.1016/j.ijrobp.2007.11.019

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  3 in total

1.  Breast cancer and funnel chest. Comparing helical tomotherapy and three-dimensional conformal radiotherapy with regard to the shape of pectus excavatum.

Authors:  M Uhl; F Sterzing; G Habl; K Schubert; H Holger; J Debus; K Herfarth
Journal:  Strahlenther Onkol       Date:  2012-01-06       Impact factor: 3.621

2.  Accelerated partial breast irradiation: what is dosimetric effect of advanced technology approaches?

Authors:  Jean M Moran; Merav A Ben-David; Robin B Marsh; James M Balter; Kent A Griffith; James A Hayman; Lori J Pierce
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-06-18       Impact factor: 7.038

3.  Tumor bed segmentation: first step for partial breast irradiation.

Authors:  Amparo González Sanchis; Luis Brualla González; Carlos Fuster Diana; Jose Carlos Gordo Partearroyo; Andres Garcia-Vilanova Comas; Jose Luis Lopez Torrecilla; Joan Roselló Ferrando
Journal:  Clin Transl Oncol       Date:  2012-07-25       Impact factor: 3.405