Literature DB >> 18206692

Computer-assisted bar-coding system significantly reduces clinical laboratory specimen identification errors in a pediatric oncology hospital.

Randall T Hayden1, Donna J Patterson, Dennis W Jay, Carl Cross, Pamela Dotson, Robert E Possel, Deo Kumar Srivastava, Joseph Mirro, Jerry L Shenep.   

Abstract

OBJECTIVE: To assess the ability of a bar code-based electronic positive patient and specimen identification (EPPID) system to reduce identification errors in a pediatric hospital's clinical laboratory. STUDY
DESIGN: An EPPID system was implemented at a pediatric oncology hospital to reduce errors in patient and laboratory specimen identification. The EPPID system included bar-code identifiers and handheld personal digital assistants supporting real-time order verification. System efficacy was measured in 3 consecutive 12-month time frames, corresponding to periods before, during, and immediately after full EPPID implementation.
RESULTS: A significant reduction in the median percentage of mislabeled specimens was observed in the 3-year study period. A decline from 0.03% to 0.005% (P < .001) was observed in the 12 months after full system implementation. On the basis of the pre-intervention detected error rate, it was estimated that EPPID prevented at least 62 mislabeling events during its first year of operation.
CONCLUSIONS: EPPID decreased the rate of misidentification of clinical laboratory samples. The diminution of errors observed in this study provides support for the development of national guidelines for the use of bar coding for laboratory specimens, paralleling recent recommendations for medication administration.

Entities:  

Mesh:

Year:  2007        PMID: 18206692     DOI: 10.1016/j.jpeds.2007.08.021

Source DB:  PubMed          Journal:  J Pediatr        ISSN: 0022-3476            Impact factor:   4.406


  4 in total

Review 1.  Effectiveness of barcoding for reducing patient specimen and laboratory testing identification errors: a Laboratory Medicine Best Practices systematic review and meta-analysis.

Authors:  Susan R Snyder; Alessandra M Favoretto; James H Derzon; Robert H Christenson; Stephen E Kahn; Colleen S Shaw; Rich Ann Baetz; Diana Mass; Corinne R Fantz; Stephen S Raab; Milenko J Tanasijevic; Edward B Liebow
Journal:  Clin Biochem       Date:  2012-06-28       Impact factor: 3.281

2.  The Efficacy of Patients' Wristband Bar-code on Prevention of Medical Errors: A Meta-analysis Study.

Authors:  M Khammarnia; A Kassani; M Eslahi
Journal:  Appl Clin Inform       Date:  2015-12-09       Impact factor: 2.342

3.  Evaluation and implementation of QR Code Identity Tag system for Healthcare in Turkey.

Authors:  Vassilya Uzun; Sami Bilgin
Journal:  Springerplus       Date:  2016-08-30

4.  Reduction in Hospital-Wide Clinical Laboratory Specimen Identification Errors following Process Interventions: A 10-Year Retrospective Observational Study.

Authors:  Hsiao-Chen Ning; Chia-Ni Lin; Daniel Tsun-Yee Chiu; Yung-Ta Chang; Chiao-Ni Wen; Shu-Yu Peng; Tsung-Lan Chu; Hsin-Ming Yu; Tsu-Lan Wu
Journal:  PLoS One       Date:  2016-08-05       Impact factor: 3.240

  4 in total