| Literature DB >> 18198775 |
Michael Voor1, Valentin Antoci, Boon Kam, Craig Roberts.
Abstract
To compare commercially available hybrid external fixator systems, mechanical testing was performed on fiberglass composite tibias with proximal metaphyseal fractures fixed with 4 hybrid systems (Ace-Fischer, Hoffmann II, Synthes Hybrid, and EBI DynaFix) and a custom "rigid" frame. The only significant difference between commercial systems was in axial loading where the Ace-Fischer allowed less motion than the other frames. In all cases the "rigid" reference frame allowed significantly less motion than the commercial systems. Approximately 50% of the motion comes from frame deformation and 50% from deformation of the pins and wires.Entities:
Mesh:
Year: 2007 PMID: 18198775 DOI: 10.3928/01477447-20071201-06