Literature DB >> 1819466

Effectiveness of palladium matrix modification for the determination of thallium by graphite furnace atomic absorption spectrometry.

Q Yang1, J Smeyers-Verbeke.   

Abstract

The effect of palladium matrix modification on the determination of thallium has been investigated. The effect of palladium combined either with ammonium nitrate or with magnesium nitrate on the sodium chloride interference was systematically studied and was compared with that of the classical sulfuric acid modifier. A combination of 6 micrograms palladium with 100 micrograms ammonium nitrate allows the direct determination of thallium in ten-fold diluted blood against matrix-free standards. The detection limit (3 sigma) in the undiluted blood was calculated to be 25 micrograms.1(-1). For the urine analysis the standard addition technique is recommended because a complete thallium recovery for all different urine samples studied was not always obtained.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1819466     DOI: 10.1016/0009-8981(91)90213-v

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  1 in total

1.  Validation of Analytical Method for Determination of Thallium in Rodent Plasma and Tissues by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS).

Authors:  James M Harrington; Eric P Poitras; Frank X Weber; Reshan A Fernando; Chamindu Liyanapatirana; Veronica G Robinson; Keith E Levine; Suramya Waidyanatha