Literature DB >> 18192656

Evaluation of a disposable device for the measurement of haemoglobin A1c in dogs.

B Catchpole1, S Mountford, S Barabas, R J Scaramuzzi.   

Abstract

A disposable device designed for measuring glycated haemoglobin (hba1c) in human blood was evaluated for use in dogs. edta blood samples were collected from 50 normoglycaemic dogs, 10 dogs suffering from anaemia and 112 diabetic dogs. hba1c was measured in all the dogs except for five of the diabetic animals, in which the concentrations were above the range of the device, that is, more than 13 per cent, and two of the anaemic dogs, in which they were below its limit of detection, that is less than 3 per cent. The diabetic dogs had higher hba1c values (range 4.9 to >13 per cent, median 9.3 per cent) than the normoglycaemic dogs (range 3.7 to 5.6 per cent, median 4.7 per cent). In the anaemic dogs the values were significantly lower (range <3.0 per cent to 5.2 per cent, median 3.5 per cent) than in the normoglycaemic dogs. There was a good correlation (R(2)=0.48) between the measurements obtained with the device and the measurements obtained with a system already validated for use in dogs.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18192656     DOI: 10.1136/vr.162.2.47

Source DB:  PubMed          Journal:  Vet Rec        ISSN: 0042-4900            Impact factor:   2.695


  1 in total

1.  Evaluation of a human glycated hemoglobin test in canine diabetes mellitus.

Authors:  Na-Yon Kim; Jaehoon An; Jae-Kyung Jeong; Sumin Ji; Sung-Hyun Hwang; Hong-Seok Lee; Myung-Chul Kim; Hyun-Wook Kim; Sungho Won; Yongbaek Kim
Journal:  J Vet Diagn Invest       Date:  2019-02-18       Impact factor: 1.279

  1 in total