Atrial lead location at the Bachmann's bundle region results in a low incidence of far field R-wave sensing.

Ewa Lewicka-Nowak1, Andrzej Kutarski, Alicja Dabrowska-Kugacka, Piotr Rucinski, Pawel Zagozdzon, Grzegorz Raczak.   

Abstract

AIMS: The aim of the study was to investigate far field R-wave sensing (FFRS) rate and characteristics at different right atrial (RA) positions in patients treated with multisite atrial pacing, with the RA lead implanted at the Bachmann's bundle (BB) area in 69 patients, in comparison to RA appendage (RAA) in 70 patients. METHODS AND
RESULTS: All measurements were done during sinus rhythm in supine patients, with unipolar (UP) and bipolar (BP) sensing configuration. The presence, amplitude threshold (FFRS trsh) and FFRS timing were determined. Sensing safety margin was defined as the ratio of sensed P-wave vs. FFRS trsh, for both the minimal (Pmin) and the mean (Pmean) P-wave amplitude. At both atrial locations BP sensing was superior to UP in FFRS rejection (P < 0.0001). At 0.5 mV sensitivity level (BP) FFRS occurred in 1% of patients at the BB site vs. 11% at the RAA (P = 0.01). FFRS trsh (BP) was 0.2 +/- 0.1 mV at the BB vs. 0.4 +/- 0.3 mV in the RAA position (P < 0.0001). Sensing safety margin, when determined for the Pmin amplitude was > or =5 in 99% of patients from the BB group, in comparison to 66% of RAA patients (P < 0.0001), in whom it was <2 in 13%. Even with the use of BP leads equipped with a 10 mm tip-to-ring spacing FFRS incidence was lower at the BB site (P < 0.01), FFRS trsh was lower (P < 0.001), and sensing safety margin was higher vs. RAA (P = 0.002).
CONCLUSION: Bachmann's bundle area features optimal conditions for signal sensing, and such atrial lead positioning may offer advantages to prevent oversensing of R-wave, thus improving functioning of standard dual chamber pacemakers, ICDs and CRT-Ds.

Entities:  

Mesh:

Year:  2008        PMID: 18192341     DOI: 10.1093/europace/eum277

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  4 in total

Review 1.  [Bachmann bundle pacing].

Authors:  C W Israel; T Burmistrava; C Berger
Journal:  Herz       Date:  2018-11       Impact factor: 1.443

2.  Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave.

Authors:  Christof Kolb; Georg Nölker; Carsten Lennerz; Hansmartin Jetter; Verena Semmler; Klaus Pürner; Klaus-Jürgen Gutleben; Tilko Reents; Klaus Lang; Ulrich Lotze
Journal:  PLoS One       Date:  2012-06-22       Impact factor: 3.240

3.  Far field R-wave sensing in Myotonic Dystrophy type 1: right atrial appendage versus Bachmann's bundle region lead placement.

Authors:  Vincenzo Russo; Gerardo Nigro; Andrea Antonio Papa; Anna Rago; Federica Di Meo; Anna Cristiano; Antonio Molino; Raffaele Calabrò; Maria Giovanna Russo; Luisa Politano
Journal:  Acta Myol       Date:  2014-10

4.  Cardiac pacing and defibrillation in children and young adults.

Authors:  Harinder R Singh; Anjan S Batra; Seshadri Balaji
Journal:  Indian Pacing Electrophysiol J       Date:  2013-01-01