Comparison of the TSI Model 8520 and Grimm Series 1.108 portable aerosol instruments used to monitor particulate matter in an iron foundry.

Yu-Hsiang Cheng1.   

Abstract

This study uses two real-time dust monitors, the TSI Model 8520 DustTrak and Grimm Series 1.108 Aerosol Spectrometer, to determine PM(10) and PM(2.5) levels simultaneously in an iron foundry. The SA Model 241 Dichotomous Sampler was used as a reference gravimetric method for comparing the measurement results obtained by these direct-reading instruments. The response to PM levels from DustTrak is higher than that of the Aerosol Spectrometer. The DustTrak provides an overestimation PM levels, and PM levels measured by an Aerosol Spectrometer are lower than actual concentrations. Calibration factors of the DustTrak and Aerosol Spectrometer are 0.74 and 1.33, respectively, when used to measure particulate matter at an iron foundry. Based on measurement results, the DustTrak provides a lower overestimation of PM(10) levels than PM(2.5) levels; that is, the response of the DustTrak increases as particle size decreases. In addition, measurement results suggest that the Aerosol Spectrometer provides precise measurements of PM(10) and PM(2.5), and measurement accuracy compared with the reference gravimetric method can be improved through a calibration factor.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18188737     DOI: 10.1080/15459620701860867

Source DB:  PubMed          Journal:  J Occup Environ Hyg        ISSN: 1545-9624            Impact factor:   2.155


  4 in total

1.  Fine and ultrafine particles emitted from laser printers as indoor air contaminants in German offices.

Authors:  Tao Tang; Julia Hurraß; Richard Gminski; Volker Mersch-Sundermann
Journal:  Environ Sci Pollut Res Int       Date:  2011-11-18       Impact factor: 4.223

2.  Evaluation of consumer monitors to measure particulate matter.

Authors:  Sinan Sousan; Kirsten Koehler; Laura Hallett; Thomas M Peters
Journal:  J Aerosol Sci       Date:  2017-02-21       Impact factor: 3.433

3.  Evaluation of the Alphasense Optical Particle Counter (OPC-N2) and the Grimm Portable Aerosol Spectrometer (PAS-1.108).

Authors:  Sinan Sousan; Kirsten Koehler; Laura Hallett; Thomas M Peters
Journal:  Aerosol Sci Technol       Date:  2016-09-07       Impact factor: 2.908

4.  A Real-Time Comparison of Four Particulate Matter Size Fractions in the Personal Breathing Zone of Paris Subway Workers: A Six-Week Prospective Study.

Authors:  Rémy Pétremand; Guillaume Suárez; Sophie Besançon; J Hugo Dil; Irina Guseva Canu
Journal:  Sustainability       Date:  2022-05-15       Impact factor: 3.889

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.