[Biological evaluation of In-Ceram-ceramics compared to cobalt-base-alloys and the metals titanium, tantalum and niobium in animal experiments].

F Limberger1, E Lenz.   

Abstract

The aim of the study was to evaluate the biocompatibility of the in-cream-ceramic-system in several technological phases by animal experiments. The comparative materials were the Co-Cr-alloys Remanium CD and Wirobond and the implant materials Titanium, Tantalum and Columbium and the high biocompatible material Teflon. The experiments were carried out in accordance with the DIN-standard 13,930 using subcutaneous implantation to verify the toxicity of the materials. 20 cylindrical test pieces were implanted into the subcutaneous connective tissue of the neck in rats. After 12 weeks the animals were sacrificed and the implants were removed with the surrounding tissues. The thickness of the fibrous capsule and the number of cells within the capsule were measured microscopically. The in-ceram-ceramic-system showed the same results like approved Co-Cr-alloys. The results of the implant materials were not attained.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1817651

Source DB:  PubMed          Journal:  Dtsch Stomatol        ISSN: 0863-4904


  1 in total

1.  Cell response to a newly developed Ti-10Ta-10Nb alloy and its sputtered nanoscale coating.

Authors:  Young-Min Kim; Mong-Sook Vang; Hong-So Yang; Sang-Won Park; Hyun-Pil Lim
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

  1 in total