| Literature DB >> 18162392 |
Nancy V Panini1, Germán A Messina, Eloy Salinas, Héctor Fernández, Julio Raba.
Abstract
This paper describes the development of an immunosensor coupled to glassy carbon electrode (GCE) modified with multiwall carbon nanotubes (MWCNT) (CNT-GCE) integrated with microfluidic systems for rapid and sensitive quantification of prostate specific antigen (PSA) in human serum samples. Mouse monoclonal (5G6) to PSA antibodies were immobilized on a rotating disk. PSA in the serum sample are allowed to react immunologically with the immobilized anti-tPSA and horseradish peroxidase (HRP) enzyme-labeled second antibodies specific to PSA. HRP, in the presence of hydrogen peroxide (H(2)O(2)) catalyzes the oxidation of 4-tert-butylcatechol (4-TBC), whose back electrochemical reduction was detected on CNT-GCE at -0.15 V. The electrochemical detection can be done within 1 min and total assay time was 30 min. The calculated detection limits for electrochemical detection and the ELISA procedure are 0.08 and 0.5 microg L(-1), respectively and the intra- and inter-assay coefficients of variation were below 4.5%. The electrochemical immunosensor showed higher sensitivity and lower time consumed than the standard spectrophotometric detection ELISA method, which shows potential for detecting PSA in clinical diagnosis.Entities:
Mesh:
Substances:
Year: 2007 PMID: 18162392 DOI: 10.1016/j.bios.2007.11.003
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618