| Literature DB >> 18155555 |
K Omoto1, K Tsuno, M Ohsaki, S Matsumura, Y Tomokiyo.
Abstract
A 1.25MV high-voltage electron microscope with a B-type omega filter has been successfully installed at Kyushu University. An image detection chamber has been set inside a concrete block below the ground level without changing the frame structure for anti-vibration. Nearly the same design as that for the 200kV microscope has been kept for the present omega filter except for its size. A new pre- and post-filter lens system with rotation-free imaging has been designed. Energy resolution, beam shape and stability of the filter have been measured. Some application data have been obtained to demonstrate the performance of the filter.Entities:
Mesh:
Year: 2007 PMID: 18155555 DOI: 10.1016/j.micron.2007.10.011
Source DB: PubMed Journal: Micron ISSN: 0968-4328 Impact factor: 2.251