[Evaluation of CoaguChek XS for measuring prothrombin time in patients receiving long-term oral anticoagulant therapy].

Jae Hyeon Lee1, Kyoung Suk Lee, Dal Sik Kim, Hye Soo Lee, Sam Im Choi, Yong Gon Cho.   

Abstract

BACKGROUND: Oral anticoagulation with warfarin requires routine monitoring of prothrombin time to maintain the international normalized ratio (INR) within the appropriate therapeutic range. Coagu- Chek XS (Roche Diagnositic, Germany) is a portable coagulometer that measures the INR. We evaluated the precision and accuracy of CoaguCheck XS by comparing it with CA-1500 (Sysmex, Japan).
METHODS: We analyzed the CV and the correlation of all INR results measured in 68 samples obtained from patients treated with warfarin and 10 samples from control subjects with no history of anticoagulant therapy with CoaguChek XS and CA-1500. We compared the turn-around time between two instruments and evaluated the differences between the results obtained with venous and capillary blood samples and those obtained with different lots of the test strip. We also evaluated the precision of the two instruments in 5 repeated tests with samples of normal and increased INR.
RESULTS: Mean INR values of 5 repeated tests with the same samples were similar. The correlation of INR values between two instruments was excellent (r2=0.97, P=0.001), and the difference in the values between the two instruments was mostly within the 95% limit of agreement, but was shown to increase in direct proportion to INR values. The turn-around time of CoaguChek XS was shorter than that of CA-1500. The differences between venous and capillary blood and between different lots of the test trip were not significant (P>0.05).
CONCLUSIONS: CoaguChek XS showed a good precision and correlation with CA-1500 with a very short turn-around time. This instrument should be clinically useful in monitoring INR of patients with oral anticoagulation.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18094572     DOI: 10.3343/kjlm.2007.27.3.177

Source DB:  PubMed          Journal:  Korean J Lab Med        ISSN: 1598-6535


  2 in total

Review 1.  [Portable coagulometer devices in the monitoring and control of oral anticoagulation therapy: a systematic review].

Authors:  Javier Caballero-Villarraso; Román Villegas-Portero; Fernando Rodríguez-Cantalejo
Journal:  Aten Primaria       Date:  2010-10-30       Impact factor: 1.137

2.  Security and cost comparison of INR self-testing and conventional hospital INR testing in patients with mechanical heart valve replacement.

Authors:  Qiu-lin Chen; Li Dong; Yi-jun Dong; Shu-lin Zhao; Bo Fu; Yu-qing Wang; Hong Jiang
Journal:  J Cardiothorac Surg       Date:  2015-01-16       Impact factor: 1.637

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.