Literature DB >> 18091251

Magnetic resonance imaging-directed method for functional neurosurgery using implantable guide tubes.

Nikunj K Patel1, Puneet Plaha, Steven S Gill.   

Abstract

OBJECTIVE: We present a magnetic resonance imaging-directed stereotactic system using implantable guide tubes for targeting deep brain nuclei in functional neurosurgery.
METHODS: Our method relies on visualization of the deep brain nuclei on high-resolution magnetic resonance images that delineate the target boundaries and enable direct targeting of specific regions of the nucleus. The delivery system comprises a modified stereoguide capable of delivering an implantable guide tube to the vicinity of the desired target. The guide tube (in-house investigational device) has a hub at its proximal end that is fixed within a burr hole and accommodates a radioopaque stylette that is inserted such that its distal end is at the desired target. After perioperative radiological confirmation of the stylette's relationship to the desired brain target, it is withdrawn from the guide tube, which may then act as a port for the implantation of an electrode for deep brain stimulation (DBS) or radiofrequency lesioning. Alternatively, the guide tube can be used to insert a catheter for drug delivery, cell transplantation, or viral-vector delivery. Implantation and verification are guided by magnetic resonance imaging or computed tomography, which enable the entire procedure to be performed under general anesthesia. The technique of implantation helps ensure optimal accuracy, and we have successfully used this device for implanting electrodes for DBS in the treatment of Parkinson's disease, essential tremor, and dystonia, and for implanting catheters for continuous delivery of glial-derived neurotrophic factor in the treatment of Parkinson's disease. The device also aids in securely fixing the DBS electrode or catheter to the cranium with ease, limiting hardware problems.
RESULTS: A total of 205 guide tubes have been implanted in 101 patients. Major complications in these cases were limited to 4% of patients. At the initial implantations, 96.3% of the guide tubes were within 1.5 mm of the target. Ten guide tubes required reimplantation secondary to target errors. With corrections, the DBS electrode was delivered to within 1.5 mm from the planned target in all cases.
CONCLUSION: This system provides a safe and accurate magnetic resonance imaging-directed system for targeting deep brain nuclei in functional neurosurgery under general anesthesia and avoids the need for electrophysiological monitoring.

Entities:  

Mesh:

Year:  2007        PMID: 18091251     DOI: 10.1227/01.neu.0000303994.89773.01

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  14 in total

Review 1.  Neuroimaging Technological Advancements for Targeting in Functional Neurosurgery.

Authors:  Alexandre Boutet; Robert Gramer; Christopher J Steele; Gavin J B Elias; Jürgen Germann; Ricardo Maciel; Walter Kucharczyk; Ludvic Zrinzo; Andres M Lozano; Alfonso Fasano
Journal:  Curr Neurol Neurosci Rep       Date:  2019-05-30       Impact factor: 5.081

2.  Deep Brain Stimulation and Swimming Performance: A Randomized Within-Person Crossover Study.

Authors:  Sarah Katherine Morgan; Omar Khan Bangash; Nat Benjanuvatra; Megan Thorburn; Irné Du Plessis; Angela Jacques; Georgina Powers; Christopher R P Lind
Journal:  Neurol Clin Pract       Date:  2021-10

Review 3.  Surgical treatment of Parkinson's disease: patients, targets, devices, and approaches.

Authors:  Aparna Wagle Shukla; Michael Scott Okun
Journal:  Neurotherapeutics       Date:  2014-01       Impact factor: 7.620

4.  Intraoperative MRI for deep brain stimulation lead placement in Parkinson's disease: 1 year motor and neuropsychological outcomes.

Authors:  Christos Sidiropoulos; Richard Rammo; Brad Merker; Abhimanyu Mahajan; Peter LeWitt; Patricia Kaminski; Melissa Womble; Adrianna Zec; Danette Taylor; Julia Wall; Jason M Schwalb
Journal:  J Neurol       Date:  2016-04-28       Impact factor: 4.849

5.  The anatomy of the caudal zona incerta in rodents and primates.

Authors:  Charles Watson; Christopher R P Lind; Meghan G Thomas
Journal:  J Anat       Date:  2013-10-21       Impact factor: 2.610

Review 6.  Deep brain stimulation for the treatment of resistant hypertension.

Authors:  Erin L O'Callaghan; Fiona D McBryde; Amy E Burchell; Laura E K Ratcliffe; Liviu Nicolae; Ivor Gillbe; Derek Carr; Emma C Hart; Angus K Nightingale; Nikunj K Patel; Julian F R Paton
Journal:  Curr Hypertens Rep       Date:  2014-11       Impact factor: 5.369

7.  Pitfalls in precision stereotactic surgery.

Authors:  Ludvic Zrinzo
Journal:  Surg Neurol Int       Date:  2012-01-14

8.  Analgesia in conjunction with normalisation of thermal sensation following deep brain stimulation for central post-stroke pain.

Authors:  Anthony E Pickering; Simon R Thornton; Sarah J Love-Jones; Charlotte Steeds; Nikunj K Patel
Journal:  Pain       Date:  2009-10-14       Impact factor: 6.961

9.  Distinct roles of dopamine and subthalamic nucleus in learning and probabilistic decision making.

Authors:  Elizabeth J Coulthard; Rafal Bogacz; Shazia Javed; Lucy K Mooney; Gillian Murphy; Sophie Keeley; Alan L Whone
Journal:  Brain       Date:  2012-10-31       Impact factor: 13.501

10.  The Effectiveness of the Stereotactic Burr Hole Technique for Deep Brain Stimulation.

Authors:  Keisuke Toyoda; Eiichirou Urasaki; Tetsuya Umeno; Waka Sakai; Akiko Nagaishi; Shunya Nakane; Takayasu Fukudome; Yuzo Yamakawa
Journal:  Neurol Med Chir (Tokyo)       Date:  2015-09-04       Impact factor: 1.742

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.