Dynamic skin cooling with an environmentally compatible alternative cryogen during laser surgery.

Wangcun Jia1, Lars O Svaasand, Thang B Nguyen, J Stuart Nelson.   

Abstract

BACKGROUND AND OBJECTIVES: Cryogen spray cooling with tetrafluoroethane (R134a) has been used to enhance epidermal protection during dermatologic laser surgery. However, R134a has a very high global warming potential (GWP = 1300). Our objective was to evaluate the cooling effectiveness of an alternative cryogen with a much lower GWP, namely liquid carbon dioxide (CO(2), GWP = 1). STUDY DESIGN/
MATERIALS AND METHODS: A thin-film thermocouple deposited on an epoxy skin phantom was used to measure surface temperature (T(s)) variations induced by R134a or CO(2) sprays. The temperature distribution in the skin phantom was estimated using T(s) and Duhamel's method. Impact pressure and noise level of both cryogen sprays were measured with a dynamic sensor and sound meter, respectively. Consumption of both cryogens was also evaluated.
RESULTS: For R134a sprays, T(s) was kept almost constant after 15 milliseconds. For CO(2) sprays, T(s) decreased continuously during the entire spurt of 50 milliseconds. The minimum T(s) induced by the CO(2) sprays was lower than that induced by R134a when the spurt duration was longer than 35 milliseconds. Numerical simulation shows that CO(2) sprays were able to induce very similar temperature reductions in the skin phantom as compared to R134a sprays when the spurt duration and delay time were selected appropriately. R134a sprays induced an impact pressure of 3.6 kPa, as compared to 43.1 kPa for CO(2) sprays. The maximum noise level for R134a sprays was 109 dBA as compared to 135 dBA for the CO(2) sprays. The R134a consumption for a 50 milliseconds spurt is 67 mg as compared to 225 mg for a CO(2) spurt of the same duration.
CONCLUSIONS: CO(2) sprays are expected to have similar skin cooling efficacy as R134a sprays. Although the CO(2) consumption is higher than R134a, its contribution to global warming is still much less than R134a. The effects of varying spurt durations on in vivo human skin and the impact on cutaneous blood flow require further investigation. (c) 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18081065     DOI: 10.1002/lsm.20582

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  2 in total

1.  In vivo laser cartilage reshaping with carbon dioxide spray cooling in a rabbit ear model: a pilot study.

Authors:  Edward C Kuan; Ashley A Hamamoto; Victor Sun; Tony Nguyen; Cyrus T Manuel; Dmitry E Protsenko; Brian J F Wong; J Stuart Nelson; Wangcun Jia
Journal:  Lasers Surg Med       Date:  2014-11-18       Impact factor: 4.025

2.  Ex vivo investigations of laser auricular cartilage reshaping with carbon dioxide spray cooling in a rabbit model.

Authors:  Edward C Wu; Victor Sun; Cyrus T Manuel; Dmitriy E Protsenko; Wangcun Jia; J Stuart Nelson; Brian J F Wong
Journal:  Lasers Med Sci       Date:  2013-01-10       Impact factor: 3.161

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.