Optimizing computed tomography to reduce artifacts from titanium aneurysm clips: an in vitro study. Technical note.

Alexander C Mamourian1, Daniel J Pluta, Clifford J Eskey, Anthony L Merlis.   

Abstract

OBJECT: At many institutions digital subtraction angiography remains the standard imaging procedure for the postoperative evaluation of patients following placement of a cerebral aneurysm clip largely because of the artifacts produced by the clip on computed tomography (CT). The authors evaluated the effect of various imaging parameters on the quality of 3D reconstructions from CT scans while imaging a phantom to optimize the CT angiograms.
METHODS: Using multidetector CT scanners with submillimeter detector collimation (0.625 mm), the authors scanned a silicone phantom with attached commercial aneurysm clips. Slice thickness, reconstruction overlap, kilovolt level, milliampere level, and pitch were varied. Neuroradiologists, who were blinded to the scanning parameters, rated the reconstructions for image quality and artifact reduction.
RESULTS: Images of the titanium clip using 140 kV and 380 mA with 0.625-mm overlapping reconstructed slices provided excellent 3D visualization of both the clip and the aneurysm model, even when using two adjacent clips.
CONCLUSIONS: Overlapping reconstructions combined with thin-section acquisition can provide detailed images of titanium clips and surrounding tissues without the use of low-pitch values.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18077966     DOI: 10.3171/JNS-07/12/1238

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  3 in total

1.  Impact of metal artifact reduction software on image quality of gemstone spectral imaging dual-energy cerebral CT angiography after intracranial aneurysm clipping.

Authors:  Vincent Dunet; Martine Bernasconi; Steven David Hajdu; Reto Antoine Meuli; Roy Thomas Daniel; Jean-Baptiste Zerlauth
Journal:  Neuroradiology       Date:  2017-07-27       Impact factor: 2.804

2.  Quantitative analysis of factors affecting cobalt alloy clip artifacts in computed tomography.

Authors:  Sook Young Sim; Chi Hoon Choi
Journal:  J Korean Neurosurg Soc       Date:  2014-11-30

3.  Gated multidetector computed tomography. A technique to reduce intracranial aneurysm clip and coil artifacts.

Authors:  A Kovács; S Flacke; H Tschampa; D Hadizadeh; S Greschus; H Clusmann; R Kristof; H Urbach
Journal:  Clin Neuroradiol       Date:  2010-05-21       Impact factor: 3.649

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.