Literature DB >> 18071377

Analysis of a scanning pentaprism system for measurements of large flat mirrors.

Julius Yellowhair1, James H Burge.   

Abstract

The optical surface of a large optical flat can be measured using an autocollimator and scanning pentaprism system. The autocollimator measures the slope difference between a point on the mirror and a reference point. Such a system was built and previously operated at the University of Arizona. We discuss refinements that were made to the hardware, the alignment procedure, and the error analysis. The improved system was demonstrated with a 1.6 m flat mirror, which was measured to be flat to 12 nm rms. The uncertainty in the measurement is only 9 nm rms.

Year:  2007        PMID: 18071377     DOI: 10.1364/ao.46.008466

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  2 in total

1.  Relaxation time and viscosity of fused silica glass at room temperature.

Authors:  M Vannoni; A Sordini; G Molesini
Journal:  Eur Phys J E Soft Matter       Date:  2011-09-23       Impact factor: 1.890

2.  Multi-beam array stitching method based on scanning Hartmann for imaging quality evaluation of large space telescopes.

Authors:  Haisong Wei; Haixiang Hu; Feng Yan; Xindong Chen; Qiang Cheng; Donglin Xue; Xuejun Zhang
Journal:  Sci Rep       Date:  2018-05-08       Impact factor: 4.379

  2 in total